84
10.5 Pneumatic Specifications
10.5.1 Pneumatic layout
1
Legend
1
Filter
20
Proximal Flow Sensor
3
150 psi Pressure Sensor
21
Linear actuator
5
45 psi pressure regulated
22
Exp pressure measurement point
6
200 LPM Proportional
23
Distal flow sensor
7
Flow Air System
2
24
HEPA filter
8
One-way valve (anti-asphyxiation)
25
x-1 / 6 psi Solenoid valve
9
Solenoid Valve On/Off (NIF)
26
Measurement point (proximal pressure differential)
10
x-5 100 psi Solenoid Valve (Nebulizer)
27
Respiratory circuit
15
O
2
concentration cell
28
HME filter
16
Internal Flow Sensor
29
Integrated valve with diaphragm
17
One-way valve
– low pressure (insp output)
30
Humidifier (optional)
18
Insp pressure measurement point
31
Nebulizer (optional)
19
On/Off Overpressure Valve
32
Capnograph (optional)
Figure 1
– Pneumatic layout
1
Pneumatic layout for Fleximag Max 700 model
2
FlowAir system only for Fleximag 300 and Fleximag 700 models