86
10.7 Distal Flow Sensor Specifications
Table 17 - Distal Flow Sensor - General Specifications
General specifications
Intended use
Measure the flow expired by the patient.
Working principle
Thermal sensor or hot wire anemometer, thermistor.
Reading range
0 to 160 SLPM
Tolerance
8%
Resistance
< 2.5mbar
Pressure range
100mbar
Lifetime
While calibration is successful
Material
MABS
10.8 Proximal Flow Sensor Specifications
Table 18 - Proximal Flow Sensor - General Specifications ADULT
General specifications
Intended use
Measure the inspired and expired flow of the patient.
Working principle
Pressure differential
Reading range
-180 to 180 SLPM
Tolerance
10%
Material
PSU
10.9 Pressure Sensor Specifications
Table 19 - Pressure sensor - General specifications
General specifications
Intended use
Measure the inspired pressure of the patient
Working principle
Pressure differential
Reading range
-60 to 120 cmH
2
O
Tolerance
5% (0 to 85°C)