6
(E)
In the standard specifications, the minimum phase difference is fixed at 50 ns for the DK800A
and 100 ns for the DK800B, but the minimum phase differences in the table below are
available as special specifications.
5-2. Output Signal Alarm
If the response speed is exceeded, the phase A/B output from this measuring unit becomes
high impedance for about 400 ms to serve as an alarm.
5-3. Receiver
*
If extending the cable, the supply voltage is +5 V±5%.
*
Use the CE22 series extension cables for bare wires (optional accessories).
A/B minimum
phase
difference
50 ns
100 ns
300 ns
500 ns
Phase A
cycle
200 ns
400 ns
1.2 µs
2 µs
Counter
allowable
frequency
5 MHz
2.5 MHz
833 kHz
500 kHz
Maximum response speed
Resolution
Resolution
0.1 µm
0.5 µm
80 m/min
250 m/min
42 m/min
100 m/min
14 m/min
33 m/min
8.4 m/min
20 m/min
Remarks
DK800A standard product
DK800B standard product
Special specifications
Special specifications
Alarm section
Phase A/B is high impedance
Line driver
Output: phase A/B/Z
Receiver
Line receiver
AM26C32 or equivalent
20 m/65.6' or less
Summary of Contents for DK805A
Page 6: ... 4 J ...
Page 8: ...ii J ...
Page 20: ... 4 CS ...
Page 22: ...ii CS ...
Page 34: ... 4 E ...
Page 36: ...ii E ...
Page 48: ... 4 G ...
Page 50: ...ii G ...
Page 59: ...DK805AFR DK805AFR5 DK805AFLR DK805AFLR5 DK805BFR DK805BFR5 DK805BFLR DK805BFLR5 ...
Page 60: ...DK805AFR DK805AFR5 DK805AFLR DK805AFLR5 DK805BFR DK805BFR5 DK805BFLR DK805BFLR5 ...
Page 62: ...DK805AFR DK805AFR5 DK805AFLR DK805AFLR5 DK805BFR DK805BFR5 DK805BFLR DK805BFLR5 ...