10
(E)
SQ57-Z Series, CH22 Series
6. Apply a thread-locking agent to the threaded areas, and install the sensor head.
Note
Take care that the reading part of the sensor head does not interfere with the scale.
At this point it is recommended to insert a 0.185 mm clearance gauge, loosely turn the mounting
screws, and check the clearance between the scale and the reading part of the sensor head.
M6 × 25
(9 N
t
m)
M6 × 25
(9 N
t
m)
M6 × 25
(9 N
t
m)
M6 × 25
(9 N
t
m)
Sensor head
Sensor head
Thread-locking
agent
Thread-locking
agent
<When using the stop surfaces>
<When using parallel pins>
Unit: mm
Fig. 3-8
7. Use the supplied clearance gauge and check that the clearance between the scale and the reading part
of the sensor head is 0.185
+ 0.065
–
0.085
mm.
* Three types of clearance gauges (0.1 mm, 0.185 mm, and 0.25 mm) are supplied. After adjusting the clearance with the
0.185 mm gauge, make sure that the 0.1 mm gauge can be inserted and the 0.25 mm cannot.
0.185
Clearance gauge
Clearance gauge
+ 0.065
– 0.085
<When using the stop surfaces>
<When using parallel pins>
Unit: mm
Fig. 3-9
Summary of Contents for SmartSCALE SQ57
Page 3: ...SQ57 Z Series CH22 Series J 1...
Page 4: ...SQ57 Z Series CH22 Series 2 J...
Page 6: ...SQ57 Z Series CH22 Series ii J...
Page 10: ...SQ57 Z Series CH22 Series 4 J 1 6 2006 42 EC...
Page 12: ...SQ57 Z Series CH22 Series 6 J 3 7 3 1 CH22 SQ57 M6 M4 M6 M8 B A 3 2 L 3 1 3 1 7...
Page 20: ...SQ57 Z Series CH22 Series 14 J 3 3 5 mm 1 0 N m 242 2 3 1 1 0 N m 3 14 1 0 N m O SQ57 Z O 3 15...
Page 23: ...SQ57 Z Series CH22 Series J 17 5 SIEMENS DRIVE CLiQ 3 5 DRIVE CLiQ...
Page 28: ...SQ57 Z Series CH22 Series 22 J 8 3 2...
Page 32: ...ii E SQ57 Z Series CH22 Series...
Page 58: ...ii G SQ57 Z Series CH22 Series...
Page 82: ......
Page 83: ...1 SQ57 Z Series CH22 Series...