10
(E)
SR27A-AZ Series, CH22 Series
2. Adjust the parallelism with respect to the machine guide of the scale rear while measuring the scale
rear with a dial gauge. Tighten the set screws.
<Measurement method>
Measure at two points where the measurement distance is as wide as possible and at the same distance
from the right and left scale ends. If using a scale with intermediate foot plate, be sure to also measure the
intermediate foot plate.
Same
distan
c
e
from
s
c
a
l
e
ends
0.
1
MG
Fig. 3-
9
3. Insert the supplied spacer (t = 1.0 mm) in the space between the scale and the scale head, and then
adjust the scale head position while bringing the scale head into contact with the scale.
Spacer
(t = 1.0 mm)
Fig. 3-10
Check that the gap at the two ends is 1 mm using a spacer (with a 1.0 mm thickness).
Unit: mm
Summary of Contents for SR27A-AZ Series
Page 3: ...SR27A AZ Series CH22 Series J 1...
Page 4: ...SR27A AZ Series CH22 Series 2 J...
Page 6: ...ii J SR27A AZ Series CH22 Series...
Page 10: ...4 J SR27A AZ Series CH22 Series 1 6 2006 42 EC...
Page 11: ...J 5 SR27A AZ Series CH22 Series 2 SR27A AZ CH22 S N...
Page 12: ...6 J SR27A AZ Series CH22 Series 3 8 3 1 CH22 SR27A M4 M8 M4 B A M8 3 2 3 2 1 3 1 ML 3 2 3 2...
Page 13: ...J 7 SR27A AZ Series CH22 Series 2 3 M8 16 20 N m M4 10 2 7 N m M4 20 2 7 N m 3 3 3 4...
Page 15: ...J 9 SR27A AZ Series CH22 Series 3 3 7 2 1 0 1 MG 0 05 MG 0 0 2 0 05 3 8 mm...
Page 16: ...10 J SR27A AZ Series CH22 Series 2 2 0 1 MG 3 9 3 t 1 0 mm t 1 0 mm 3 10 t 1 0 mm 1 mm mm...
Page 17: ...J 11 SR27A AZ Series CH22 Series 3 3 5 mm 1 0 N m 242 1 0 N m 1 3 2...
Page 27: ...J 21 SR27A AZ Series CH22 Series 9 3...
Page 28: ...22 J SR27A AZ Series CH22 Series...
Page 32: ...ii E SR27A AZ Series CH22 Series...
Page 54: ...22 E SR27A AZ Series CH22 Series...
Page 58: ...ii G SR27A AZ Series CH22 Series...
Page 80: ...22 G SR27A AZ Series CH22 Series...
Page 82: ......
Page 83: ...SR27A AZ Series CH22 Series 1...