A. Optional Accessories
Foot Switch
The Foot Switch option (P/N: 09912508) provides for external control of
the
cC
harge or
mM
agnetize function
s
. Other provisions for remote operation are available.
Temperature Monitor
The Temperature Monitor option (P/N: 09003126) will continuously display the
B. Safety Procedures
This equipment operates at high voltage. All high voltage windings and connections are
enclosed in the cabinet and Magnetizing Fixture housing. Under no circumstances should
these housings or the cabinet be tampered with, or the design altered unless previously
discussed with Magnetic Instrumentation, Inc. Care should be taken that the equipment is
properly grounded. No drilling, tapping, or machine work should be done on the
Magnetizing Fixture housing unless drawings of the internal construction are first
obtained from Magnetic Instrumentation, Inc.
An interlock is mounted to the panel covering the disconnect jacks for the
Magnetizing
Fixture
. This prevents operation of the machine without the
Magnetizing Fixture
properly
attached. The high-voltage-disconnect jack and tT
he four side panels
should only be
removed are provided with interlocks that break the power line. These interlocks should
be blocked only
for troubleshooting purposes and then only by an engineer or skilled
technician familiar with this type of equipment and with the danger involved with
working with several hundred microfarads of capacitance charged to 3000 volts.
Caution labels are placed on the equipment in strategic locations, to alert the possible
safety hazards. For maximum safety to the operator, all safety precautions indicated or
implied must be observed.