5
REPLACEMENT PARTS
Replacing electronics/transducer
Replacement parts
Echotel electronics can be removed in the field under process conditions. Follow below steps to exchange electronics/trans-
ducer:
Note:
Adjust set up of the replacing electronics following the settings of the old electronics (see configuration section)
1. Disconnect power before removing the housing cover
2. Remove power/output wires (a)
3. Remove the 2 bracket screws and slide out electronics (b)
4. Remove the transducer wires (c)
5. Re-assemble following the same procedure in opposite way. Make sure that the tip on the bracket of the electronic block is
seated properly in the corresponding recess in the housing base - (d)
No.
Description
Part Number
1
Cast aluminium cover
Blind
With window
Deep drawn stainless steel cover
Blind
With window
004-9192-009
036-4410-010
036-5702-003
036-5702-002
2
“O”-Ring
012-2201-237
3
Electronic module
089-7255-001
4
Transducer
See model
identification
2
1
3
4
3
3
a
d
c
b