background image

6

CALIBRA

TION

AUXILIARY CURRENT METER

In order to calibrate the transmitter

, you may use either the

loop current meter, or a second current meter which has an
internal resistance of less than 10 Ω.
It should read currents in the range of 1 to 25 mA, with a 
resolution of 0.01 mA. Using a meter of less resolution will
somewhat reduce the calibration accuracy. 
An error of 0.2 mA is equivalent to a 1% error, based on the
full scale current of 20 mA. 

To calibrate the instrument:

1. Remove transmitter housing cover.

2. Check that the LED is glowing (shadowing the LED by

hand may be helpful in brightly lit environments).

NOTE

: The LED will glow brighter with increasing signal.

a. If the LED is lit, continue to step 3.

b. If the LED isn’t lit, check for supply voltage of 14 to

40 V at the terminal strip. Also confirm that the 
polarity is correct. If LED still does not light, replace
the cover and contact factory.

3. Connect the auxiliary meter’s positive lead to Test point

1 and its negative lead to Test point 2. See 

Figure 7

.

NOTE

: This use of TP- and TP+ allows the loop current to

be monitored at the location where the adjustments are
being made, without removing any wires. 

It is not necessary to use this auxiliary meter if another
means is available to measure loop current.

a. If the LED goes completely out, proceed to step 4.

b. If the LED doesn’t go out, check the following:

• the meter is properly connected.
• the appropriate meter range has been selected.
• meter resistance is less than 10 Ω for the 

selected range.

4. The loop current should now be in the range of 1.5 mA

to 38 mA, which is normal at this point.

TWO POINT CALIBRA

TION - LEVEL INCREASE

T

wo calibration procedures are described on page 7. Follow

the one which fits your application.
Please note the following definitions used in the calibration
procedures, referring to 

Figure 8

.

L0 =

the level of material in the vessel which corresponds
to 4.0 mA of loop current, i.e., the 0% level.

L1 =

a material level higher than L0.

L2 =

a material level higher than L1, but less than L3.

L3 =

the level of material in the vessel which corresponds
to 20.0 mA of loop current, i.e., the 100% level.

NOTE

: T

o avoid the possibility of a “dead zone”, L0 must 

be at least 50 mm (2") above the end of the probe for 
conductive media and 100 mm (4") above for non-
conductive media.
Calibration A = when material in tank can be set to L0 (0%)

and L3 (100%).

Calibration B = when material in tank can be set to L1

(greater than 0%) and L2 (less than 100%).

NOTE

: Calibration Procedure A gives the most accurate

results and is the recommended procedure in all cases.

+

TP +

-

TP -

B

A

Figure 8

Figure 7

Auxiliary

current

meter

L3 - 100 %

L2 - less than 100 %

L1 - greater than 0 %

L0 - 0 %

Calibration

procedure

Summary of Contents for Kotron RF 2-Wire Series

Page 1: ...dent upon the probe s total surface area Adjusting the combination of the probe s diameter and length and of course its proximity to ground in any given application can generate the necessary capacita...

Page 2: ...elect code 0 or 4 in housing selections HOUSING IP66 OUTPUT SIGNAL 3 4 to 20 mA output direct action 4 20 to 4 mA output reverse action 0 3 4 NPT cable entry aluminium 1 NPT with local indicator 1 M20...

Page 3: ...tory In all cases a good electrical connection must be made between the ground surface and the probe housing MOUNTING Two wire transmitters with probes up to and including 305 mm 12 in length are ship...

Page 4: ...shing off probe 9 While holding probe cable loosen upper packing gland nut 10 Pull excess cable up through packing gland nut until cable is taut 11 Tighten the packing gland nut 12 Cut off cable 35 mm...

Page 5: ...r bracket The meter is attached by two wires to the terminal block underneath it Pull meter gently out of the housing and lay the meter to one side being careful not to disturb the wiring See Figure 6...

Page 6: ...ailable to measure loop current a If the LED goes completely out proceed to step 4 b If the LED doesn t go out check the following the meter is properly connected the appropriate meter range has been...

Page 7: ...until the loop current is exactly L2mA Calibration is now complete CALIBRATION PROCEDURE A 1 Move the material in the vessel to its L0 0 level 2 Turn the zero coarse zero ne and the ne span calibratio...

Page 8: ...n controls incorrectly set Turn span potentiometers fully clockwise before calibrating zero c Excessive probe capacitance Maximum zeroing range is 450 pf zero dip switch 1 in closed position 650 pf ze...

Page 9: ...chart 1 on page 5 Reduce loop resistance or Increase supply voltage consult factory 4 Loop current less than 4 mA a Incorrect calibration Recalibrate unit b Material level below 0 No corrective actio...

Page 10: ...ard Base Mounting bracket Pipe nipple Bushing O ring Cover Cover assembly O ring Meter assembly Base Description Speci cation Supply Voltage 14 to 40 V DC Current 38 mA max Line Variation Less than 0...

Page 11: ...ptional Mounting Flange Optional Mounting Flange 10 375 13 50 16 625 10 375 13 50 16 625 321 12 65 118 4 63 76 3 00 Probe 102 4 00 10 375 127 5 00 13 50 16 625 Probe 102 4 00 10 375 127 5 00 13 50 16...

Page 12: ...it is essential that a Return Material Authorisation RMA form will be obtained from the factory It is mandatory that this form will be attached to each material returned This form is avail able throug...

Reviews: