Probe
high
Probe low
Figure 14
3. Press the DOWN
arrow key to get to MODE. Press
ENT .
4. Scroll to SING. Press
ENT
. Unit reads
NOT READY
.
You are now in the Single Point Mode.
5. Return to
CNFG
and repeat the above procedure now
choose
MULT
from the
MODE
menu. Press
ENT
. Unit
reads
MAGNETROL INTL M M M…
.
6. All of the original data has now been cleared.
7. Begin
Probe Calibration
procedure again from the start of
Program Path 3
(page 17).
2.4.2.2 Error Messages in PRLO/HI Menus
1.
2BIG:
Occurs during
PRLO
or
PRHI
data entry when
displayed value exceeds 19999.
2.
2SML:
Occurs during
PRLO
data entry when displayed
value underflows 0, or during
PRHI
when displayed value
underflows 1.
3.
2 LO:
Occurs when stored
PRHI
measurement (not
display) would be less than currently stored
PRLO
measurement. In other words, the capacitance being
used to set the
PRHI
point is less than the capacitance
used to set the
PRLO
.
4.
2 HI:
Occurs when stored
PRLO
measurement (not
display) would be more than currently stored
PRHI
measurement. In other words, the capacitance being
used to set the
PRLO
point is more than the capacitance
used to set the
PRHI
point.
5.
UNDR:
Occurs when displayed value goes below -999.
6.
OVER:
Occurs when displayed value goes above 19,999
(red, >10,000 LED illuminated and 9999 on the display).
ENT
ENT
ENT
50-621 Kotron Sentinel Models 822, 832 and 842 RF Point Level Switches
19