PCCU configuration and calibration unit
Configuration Procedure
•
This procedure is the same as the PCU above.
Calibration Procedure
•
This allows for the calibration of the Soldering iron and the Hot air pencil to a
known calibrated source (Type K thermocouple simulator)
Calibrating the PCCU
•
If required this procedure can be done by the factory.
Equipment required (This section can be omitted if PCCU is pre-calibrated)
•
PCCU unit
,
Type K thermocouple simulator with input cable
•
PC with calibration program loaded from the disc supplied
•
Female DB-9 serial cable, RJ-12 to RJ-12 connector
.
Procedure
•
Connect the PCCU to the PC using the serial cable
•
Connect the thermocouple simulator to the Soldering iron input of the PCCU.
•
Load the
Calibration program
on the PC, select the correct Com port and reset the
PCCU.
•
The screen on the PC will start with a instruction to connect the simulator to the
soldering iron input (if this does not happen try another Com port, check serial cable and
PCCU)
•
Follow the instructions on the PC and
only click the OK button when the
simulator’s temperature has been set to the requested temperature and instructions
to change inputs have been carried out.
•
If, at any time the procedure is interrupted, the
Calibration will be invalid and has
to be restarted.
IT IS VERY IMPORTANT THE ABOVE PROCEDURE IS CARRIED OUT
CORRECTLY.
Calibrating the 3005 Station
This procedure automatically calibrates the 3005 Station’s Soldering iron and Hot air
pencil to the PCCU, which has been precalibrated to a certified source.