TE100 With T4000 Quick Reference Guide 

 

 

 

Revision 002 

Calibration 

Important: The following procedure assumes that all load 
cells are properly installed and that no outside forces are 
acting on or coming into contact with the scale/vessel. The 
procedure also assumes that the TE100 is turned on and has 
been allowed to warm up for a minimum of fifteen (15) 
minutes and that the T4000 has been properly configured. 

1.

 

Make sure the T4000 has been configured for the intended 
application. This includes setting the units, decimal point, 
scale capacity, averages, etc.  

2.

 

From the Summary display press the button under MENUS. 
The Configuration menu appears. 

3.

 

Press the down arrow button until the cursor is in front of 
Calibration. 

 

4.

 

Press the Enter button. The Calibration menu appears. 

5.

 

Scroll down to the Trad Cal menu line.  Press Enter. This 
method of calibration uses test weights. Taylor Products 
recommends that the test weights total 80 – 100% of the 
desired package weight, and that the weight be distributed 
uniformly on/in the scale. 

 

6.

 

Press the Enter button. The Trad Cal menu appears. 

 

7.

 

Press the Enter button. The Zero Value weight menu appears. 

8.

 

Traditional Calibration requires a zero point and the physical 
placement of test weights on the scale. To set the Zero Value 
Weight: 

a.

 

Remove all weight “live load” from the Scale. The 
Zero Value should be 0.0. 

 

CAUTION 

The scale MUST be empty. 

b.

 

Wait 12 seconds or more. 

9.

 

Use the left and right arrow buttons to position the cursor. Use 
the up or down arrow buttons to Enter the Zero Value for this 
instrument.  

Note: Most applications will set the Zero Value to 0.00. 

10.

 

Press the Enter button to save the entry. 

11.

 

Press the down arrow button until the cursor is in front of “Do 
Trad Cal ” (Zero)”. 

 

12.

 

Press the Enter button to Do Trad Cal (Zero). The controller 
may request password verification. Enter the password, press 
Enter, then select the Do Trad Cal (Zero) again. 

a.

 

A “Cal Completed OK” message appears briefly if 
the calibration was successful. 

b.

 

An Error number appears if the calibration was not 
successful. Refer to Troubleshooting in the Model 
TE100 Electronic Small Bagging Scale Operation 
and Maintenance Manual. 

13.

 

Press the down arrow button until the cursor is in front of 
Span Value. 

 

14.

 

Press the Enter button. The Span Weight menu appears. The 
last Span Weight is displayed. 

 

15.

 

To set the Span Weight: 

a.

 

Place a certified test weight on the scale. 

b.

 

Use the left and right arrows to position the cursor 
and the up or down arrows to Enter the value of the 
test weight. If a 10 lb. weight is used, Enter 10.00. 

c.

 

Press the Enter button to save the entry. 

16.

 

Press the down arrow button until the cursor is in front of “Do 
Trad Cal (Span)”. 

 

17.

 

Press the Enter button to Do Trad Cal (Span). The controller 
may request password verification. Enter the password, press 
Enter, then select the Do Trad Cal (Span) again. 

a.

 

A “Cal Completed OK” message appears briefly if 
the calibration was successful. 

b.

 

An Error number appears if the calibration was not 
successful. Refer to Troubleshooting in the Model 
TE100 Electronic Small Bagging Scale Operation 
and Maintenance Manual. 

18.

 

Press the function button under EXIT three times to return to 
the Standby Screen. 

19.

 

The scale is now calibrated. 