TE100 With T4000 Quick Reference Guide
2
Revision 002
Calibration
Important: The following procedure assumes that all load
cells are properly installed and that no outside forces are
acting on or coming into contact with the scale/vessel. The
procedure also assumes that the TE100 is turned on and has
been allowed to warm up for a minimum of fifteen (15)
minutes and that the T4000 has been properly configured.
1.
Make sure the T4000 has been configured for the intended
application. This includes setting the units, decimal point,
scale capacity, averages, etc.
2.
From the Summary display press the button under MENUS.
The Configuration menu appears.
3.
Press the down arrow button until the cursor is in front of
Calibration.
4.
Press the Enter button. The Calibration menu appears.
5.
Scroll down to the Trad Cal menu line. Press Enter. This
method of calibration uses test weights. Taylor Products
recommends that the test weights total 80 – 100% of the
desired package weight, and that the weight be distributed
uniformly on/in the scale.
6.
Press the Enter button. The Trad Cal menu appears.
7.
Press the Enter button. The Zero Value weight menu appears.
8.
Traditional Calibration requires a zero point and the physical
placement of test weights on the scale. To set the Zero Value
Weight:
a.
Remove all weight “live load” from the Scale. The
Zero Value should be 0.0.
CAUTION
The scale MUST be empty.
b.
Wait 12 seconds or more.
9.
Use the left and right arrow buttons to position the cursor. Use
the up or down arrow buttons to Enter the Zero Value for this
instrument.
Note: Most applications will set the Zero Value to 0.00.
10.
Press the Enter button to save the entry.
11.
Press the down arrow button until the cursor is in front of “Do
Trad Cal ” (Zero)”.
12.
Press the Enter button to Do Trad Cal (Zero). The controller
may request password verification. Enter the password, press
Enter, then select the Do Trad Cal (Zero) again.
a.
A “Cal Completed OK” message appears briefly if
the calibration was successful.
b.
An Error number appears if the calibration was not
successful. Refer to Troubleshooting in the Model
TE100 Electronic Small Bagging Scale Operation
and Maintenance Manual.
13.
Press the down arrow button until the cursor is in front of
Span Value.
14.
Press the Enter button. The Span Weight menu appears. The
last Span Weight is displayed.
15.
To set the Span Weight:
a.
Place a certified test weight on the scale.
b.
Use the left and right arrows to position the cursor
and the up or down arrows to Enter the value of the
test weight. If a 10 lb. weight is used, Enter 10.00.
c.
Press the Enter button to save the entry.
16.
Press the down arrow button until the cursor is in front of “Do
Trad Cal (Span)”.
17.
Press the Enter button to Do Trad Cal (Span). The controller
may request password verification. Enter the password, press
Enter, then select the Do Trad Cal (Span) again.
a.
A “Cal Completed OK” message appears briefly if
the calibration was successful.
b.
An Error number appears if the calibration was not
successful. Refer to Troubleshooting in the Model
TE100 Electronic Small Bagging Scale Operation
and Maintenance Manual.
18.
Press the function button under EXIT three times to return to
the Standby Screen.
19.
The scale is now calibrated.