Industrial Gaussmeter IGM11 Operating Instructions
Chapter 4 Control Elements and Connections
Page 21 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
4.5
USB Interface
The transfer of the measuring values and the entire gaussmeter control are
possible via the installed standard USB interface on the front panel.
The interface is specified according to the USB (Universal Serial Bus) 2.0
standard. The interface is not insulated. Please note that the shielding of
the measuring probes (e.g. the metallic handle) might be connected to the
protective conductor of your personal computer via this USB interface.
The device supports the USB device class HID. The installation of a special
driver on the computer is required.
Via the driver all functions and parameters of the device can be adjusted
using SCPI commands, and the measued values can be read out
automatically.
In seldom cases, there might be troubles when connecting the device. In
these cases, we recommend the connection via a USB hub.