5.5
Measuring Range
The gaussmeter has 4 measuring ranges. The range limit value is shown at the
bottom left area of the display. The range limit values depend on the ranges
themselves and on the selected measuring unit. The ranges can only be switched
manually. Normally, the measuring range is adjusted in the set-up menu. Regarding
the manual switch, the measuring range is selected by multiple pressing of a button
assigned in the set-up menu.
The following table shows an overview of the measuring ranges.
DC Field Measurement
AC Field Measurement
Fast Pulse Measurement
Unit
Range
Range Limit
Resolution
Range Limit
Resolution
Range Limit
Resolution
Tesla
1
10 mT
1 µT
10 mT
10 µT
10 mT
10 µT
T
2
100 mT
10 µT
100 mT
100 µT
100 mT
100 µT
3
1000 mT
100 µT
1000 mT
1 mT
1000 mT
1 mT
4
4500 mT
[10 T]
1 mT
3000 mT
[6.5 T]
10 mT
4500 mT
[10 T]
10 mT
Ampere/m
1
10 kA/m
1 A/m
10 kA/m
10 A/m
10 kA/m
10 A/m
A/m
2
100 kA/m
10 A/m
100 kA/m
100 A/m
100 kA/m
100 A/m
3
1000 kA/m
100 A/m
1000 kA/m
1 kA/m
1000 kA/m
1 kA/m
4
3800 kA/m
[8500 kA/m]
1 kA/m
2500 kA/m
[5500 kA/m]
10 kA/m
3800 kA/m
[8500 kA/m]
10 kA/m
Gauss
1
100 G
10 mG
100 G
100 mG
100 G
100 mG
G
2
1000 G
100 mG
1000 G
1 G
1000 G
1 G
3
10 kG
1 G
10 kG
10 G
10 kG
10 G
4
45 kG
[100 kG]
10 G
30 kG
[65 kG]
100 G
45 kG
[100 kG]
100 G
Oersted
1
100 Oe
10 mOe
100 Oe
100 mOe
100 Oe
100 mOe
Oe
2
1000 Oe
100 mOe
1000 Oe
1 Oe
1000 Oe
1 Oe
3
10 kOe
1 Oe
10 kOe
10 Oe
10 kOe
10 Oe
4
45 kOe
[100 kOe]
10 Oe
30 kOe
[65 Oe]
100 Oe
45 kOe
[100 kOe]
100 Oe
Figure 14 Measuring Ranges
The values in [] apply to the operation with high-field probes.