Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 66 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
:STATus:PRESet deletes the release register for the error data. When querying an
event register, the device returns a decimal value that corresponds to the sum of the
binary place values of all bits that are set in this register. For setting the bits in a
release register, a decimal value is transmitted that corresponds to the sum of the
binary place values of all bits set in this register.
9.10.4.3
Sum Register
In the sum register the results of the event registers are summarized after the
masking in the release register and assigned to single bits. The deletion of the event
in the event registers also sets back the respective sum bit in the status register. Via
the release register of the sum register the single sum bits can be re-merged to one
whole message service query.
9.10.4.4
Overview Status Model
Overload
0
Data ready
1
Not used
Not used
Not used
Not used
Not used
Not used
Event Register
Measurement
STAT:MEAS:EVEN?
Release Register
STAT:MEAS:ENAB <
Value
>
STAT:MEAS:ENAB?
OR
Not used
Calibration Error
1
Not used
Not used
Not used
Not used
Internal Cal. Error
6
Probe Cal. Error
7
Event Register
Error Data
STAT:QUES:EVEN?
Release Register
STAT:QUES:ENAB <
Value
>
STAT:QUES:ENAB?
OR
Operation ended
0
Not used
Not used
Internal Error
3
Not used
SCPI-Error
5
Not used
On
7
Event Register
Standard
*ESR?
Release Register
*ESE <
Value
>
*ESE?
OR
Meas. Event
0
Not used
Not used
Error Data
3
Not used
Standard Event
5
Service-Request
6
Not used
Sum Register
*STB?
Release Register
*SRE <
Value
>
*SRE?
OR
Trial Version
Subje
ct to chan
ge