Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 67 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.10.4.5
Definition of the Register Bits
Definition of the bits in the measurement event register
Bit
Decimal
Value
Description
Definition
0
1
Overflow
Range overflow during measurement
1
2
Data available
Measurement terminated. Data available.
Definition of the bits in the error event register
Bit
Decimal
Value
Description
Definition
1
2
General
Calibration Error
Is set if the calibration data of the measuring
probe could not be read or if the internal
calibration data are not consistent.
6
64
Internal
Calibration Error
Is set if the internal calibration data are not
consistent.
7
128
Probe
Calibration Error
Is set if the calibration data of the measuring
probe are not consistent.
Definition of the bits in the standard event register
Bit
Decimal
Value
Description
Definition
0
1
Command executed
(OPC, operation complete)
Is set at the end of the SCPI command processing
3
8
Device error
(DDE, device-dependend
error)
An error has occured.
5
32
Command error
(CME, command error)
An error has occcured during the SCPI command
processing.
7
128
Ready for use
(PON, power-on)
Is set when the device is ready for use.
Definition of the bits in the sum register
Bit
Decimal
Value
Description
Definition
0
1
Measurement event
One or several bits are set in the measurement event
register and activated in the release register.
3
8
Error event
One or several bits are set in the error event register
and activated in the release register.
5
32
Standard event
One or several bits are set in the standard event
register and activated in the release register
6
64
Service request
One or several bits are set in the sum register and
activated in the release register.
Trial Version
Subje
ct to chan
ge