Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 73 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.1.7
*OPC
Description:
Operation Complete Command)
Set the bit "Operation complete" (bit 0) in the standard event register
after the command has been completed.
Mode:
Command
Parameter:
none
Response:
none
*RST
Value:
not relevant
Example:
send
*OPC<LF>
9.12.1.8
*SRE[?]
Description:
S
ervice
R
equest
E
nable Query
Set or query the status release register. The gaussmeter returns the
bit pattern as a decimal value.
Mode:
Command and query
Parameter:
{<Value>} within the range of 0..255
Response:
Status release register
*RST
Value:
not relevant
Example:
send
*SRE 22<LF>
send
*SRE?<LF>
receive 41<CR><LF>
9.12.1.9
*STB?
Description:
Read
St
atus
B
yte Query
Query the status register.
Mode:
Query
Parameter:
none
Response:
Status register
*RST
Value:
not relevant
Example:
send
*STB?<LF>
receive 0<CR><LF>
Trial Version
Subje
ct to chan
ge