Magtrol Model 6530 Three-Phase Power Analyzer
Chapter 3 – Installation/Configuration
SETUP
3.3.3.2
Software Configuration
To configure the 6530 external sensor, complete the following steps utilizing the USER MENU
located on the front panel of the unit.
1. Turn on the 6530. See
Section 3.1 – Powering Up the 6530
.
2. Press ENTER button. The display should appear as shown in
Figure 2–3 Device Setup
Menu
.
3. Press
and
buttons until EXTERNAL SENSOR appears in the display.
4. Press ENTER button. The display will appear as follows.
CURRENT
VOLTAGE
POWER/PF
Figure 3–20 External Sensor Scale Factor Setup Menu
5. To set the scale factor for A1, press the RMS button under A1 and use the
buttons in the USER MENU to adjust the value.
6. To set the scale factor for A2, press the RMS button under A2 and use the
buttons in the USER MENU to adjust the value.
7. To set the scale factor for A3, press the AVG button under A3 and use the
buttons in the USER MENU to adjust the value.
8. Press ENTER to exit the Device Setup Menu.