Magtrol Model 6530 Three-Phase Power Analyzer
Chapter 7 – Optional Features
OPTIONS
5. To set amps value, press the RMS button under AMPS and use the
buttons in the
USER MENU to adjust the value.
6. To set volts value, press the RMS button under VOLTS and use the
buttons in the
USER MENU to adjust the value.
7. To set watts value, press the AVG button under WATTS and use the
buttons in
the USER MENU to adjust the value.
8. Press ENTER to exit the Device Setup Menu.
7.1.5
C
ALIBRATION
7.1.5.1
Closed-Box Calibration
The 6530 features closed-box calibration for the analog output. The advantage of closed-box
calibration is that the user does not have to disassemble the case or make mechanical adjustments.
7.1.5.2
Calibration Schedule
Calibrate the 6530 analog output board:
•
After any repairs are performed.
•
At least once a year; more frequently to ensure required accuracy.
7.1.5.3
Calibration Commands
The following commands are accessible when the instrument is placed in calibration mode by
holding the SHIFT key while the power is turned on. "CALIBRATION MODE ENABLED" will
appear on the display.
Note:
If the unit has not been placed in calibration mode, all calibration
commands will return "CAL DISABLED" response.
Response to all analog output calibration commands:
Default:
output_string = <NULL_STRING (0x00) ><terminator>