26 | ACX1281 | Troubleshooting
en
7.2
Refrigerant identification unit
Error code Messages
Action
D000
Sample NOK
• Move the unit away from sources of EMF or RFI such as radio
transmitters and arc welders
• Repeat refrigerant analysis
D001
Air or gas reading was unstable
• Move the unit away from sources of EMF or RFI such as radio
transmitters and arc welders
• Repeat refrigerant analysis
D002
Air or gas reading was excessively high
• Prevent refrigerant from flowing into the unit through the sample inlet during
air calibration
• Allow any refrigerant in the atmosphere to dissipate before performing air cal
-
ibration
• Repeat refrigerant analysis
D003
Air calibration resulted in low output
• Move the unit to an area where the ambient temperature is within the speci
-
fied operating range
• Repeat refrigerant analysis
D004
Unit is beyond temperature range
• Repeat refrigerant analysis
D005
Excessive air or little or no sample flow
• Verify the coupler valve is open
• Repeat refrigerant analysis
• Verify the sample filter is not plugged with debris or oil.
• Replace brass sample filter
D006
Internal error
• Retry Operation. Contact tech-support if problem persists
D007
User abort
• Repeat refrigerant analysis
D008
Communication error
• Check USB connection