Setting and measuring functions
(Please open page 24 for illustrations)
Putting into operation:
Switching-on
– press key 2.3 (ON/OFF) for less than 2 sec.
Note: The display shows the last activated mode before switching-off.
Switching-off
– press key 2.3 (ON/OFF) for more than 2 sec.
Note: Even after switching-off the last displayed value is memorized.Immediately after switching-on
measuring can commence providing that the position of the right measuring arm holder was not
changed.
Measuring function:
Changing of the measuring unit (mm/inch) (Fig. A)
– press key 2.4 for less than 2 sec.
RESET (zeroing) (Fig. B)
– press key 2.3 for less than 2 sec. The display shows either 0 or the last stored PRESET-value,
depending on the active reference level.
Reference function:
Switching to reference mode (REF I/II) (Fig. C)
– press key 2.4 for more than 2 sec.
Selecting reference mode (REFI/II)
– press key 2.4 for less than 2 sec.
HOLD-mode (Fig. D)
– press key 2.3 for less than 2 sec.
– to cancel, press key 2.3 once more for less than 2 sec.
Note: The HOLD-function is only available in REF I or REF II. When the instrument is connected to a
peripheral unit (e.g. printer) via data exit 2.5., the stored value will be accepted.
A reset into the measuring function is not possible as long as "HOLD" is displayed.
PRESET- input function:
– select REF I or II,
– press key 2.4 until PRESET appears in the display (Fig. E)
– using key 2.3 select counting direction
– using key 2.4 select the position of the first digit
– using key 2.3 set required value
– continue with key 2.4 and key 2.3 until the required value is entered
– finally for confirmation and return to the measuring function press key 2.4 for more than 2 sec.
12
Note: The instrument cannot be switched off while the PRESET-input function is active.
MIN/MAX- function:
Automatic storage and display of MIN and/or MAX as well as display of MIN-MAX (difference,
absolute).
– press key 2.4 for more than 2 sec. until MIN appears in the display (Fig. F)
Note: After release of the key 2.4 the last activated function MIN, MAX or MIN - MAX (difference,
absolute) appears. The register MIN and MAX are then refreshed with the actual value.
– to select the required function MIN, MAX or MIN - MAX subsequently press key 2.4 repeatedly for
less than 2 sec. until the required function is obtained.
Call-up of the MIN- and MAX- register:
– press key 2.4 for less than 2 sec
Tolerance function:
Exceeding the enterred high and low tolerance limit is indicated by arrows thus TOL
▲
resp.TOL
▼
Entering the high tolerance limit:
– press key 2.4 for more than 2 sec. until TOL appears in the display (Fig. G)
– followed by pressing key 2.4 for less than 2 sec. to select the high tolerance limit input function
– enter the high tolerance limit in the same manner as enterring a PRESET- value (Fig. H)
Entering the low tolerance limit:
– press key 2.4 for more than 2 sec.
– enter the low tolerance limit in the same manner as enterring a PRESET- value
Return to the tolerance function:
– press key 2.4 for more than 2 sec. until SET disappears from the display
Changing the counting direction:
– simultaneously press key 2.3 and 2.4 for more than 2 sec. The display continually changes
between 0000.01 and – 0000.01 (not possible in PRESET-function). As soon as the resolution
with the required counting direction appears release both keys.
Use with OPTO-RS 232 -plug:
The adapter plug opto-RS 232 permits the remote control of all key functions of the instrument from a
computer (PC or SQS-system).