Setting and measuring functions

(Please open page 24 for illustrations)

Putting into operation:

Switching-on
–  press key 2.3 (ON/OFF) for less than 2 sec.

Note: The display shows the last activated mode before switching-off.

Switching-off
–  press key 2.3 (ON/OFF) for more than 2 sec.

Note: Even after switching-off the last displayed value is memorized.Immediately after switching-on
measuring can commence providing that the position of the right measuring arm holder was not
changed.

Measuring function:

Changing of the measuring unit (mm/inch) (Fig. A)
–  press key 2.4 for less than 2 sec.

RESET (zeroing) (Fig. B)
–  press key 2.3 for less than 2 sec. The display shows either 0 or the last stored PRESET-value,
    depending on the active reference level.

Reference function:

Switching to reference mode (REF I/II) (Fig. C)
–  press key 2.4 for more than 2 sec.

Selecting reference mode (REFI/II)
–  press key 2.4  for less than 2 sec.

HOLD-mode (Fig. D)
–  press key 2.3 for less than 2 sec.
–  to cancel, press key 2.3 once more for less than 2 sec.

Note: The HOLD-function is only available in REF I or REF II. When the instrument is connected to a
peripheral unit (e.g. printer) via data exit 2.5., the stored value will be accepted.
A reset into the measuring function is not possible as long as "HOLD" is displayed.

PRESET- input function:

–  select REF I or II,
–  press key 2.4 until PRESET appears in the display (Fig. E)
–  using key 2.3 select counting direction
–  using key 2.4  select the position of the first digit
–  using key 2.3 set required value
–  continue with key 2.4 and key 2.3 until the required value is entered
–  finally for confirmation and return to the measuring function press key 2.4 for more than 2 sec.

12

Note: The instrument cannot be switched off while the PRESET-input function is active.

MIN/MAX- function:

Automatic storage and display of MIN and/or MAX as well as display of MIN-MAX (difference,
absolute).

– press key 2.4 for more than 2 sec. until MIN appears in the display (Fig. F)

Note: After release of the key 2.4 the last activated function MIN, MAX or MIN - MAX (difference,
absolute) appears. The register MIN and MAX are then refreshed with the actual value.

– to select the required function MIN, MAX or MIN - MAX subsequently press key 2.4 repeatedly for

less than 2 sec. until the required function is obtained.

Call-up of the MIN- and MAX- register:

– press key 2.4 for less than 2 sec

Tolerance function:

Exceeding the enterred high and low tolerance limit is indicated by arrows thus TOL 

 resp.TOL 

Entering the high tolerance limit:

– press key 2.4 for more than 2 sec. until TOL appears in the display (Fig. G)
– followed by pressing key 2.4 for less than 2 sec. to select the high tolerance limit input function
– enter the high tolerance limit in the same manner as enterring a PRESET- value (Fig. H)

Entering the low tolerance limit:

– press key 2.4 for more than 2 sec.
– enter the low tolerance limit in the same manner as enterring a PRESET- value

Return to the tolerance function:

– press key 2.4 for more than 2 sec. until SET disappears from the display

Changing the counting direction:

– simultaneously press key 2.3 and 2.4 for more than 2 sec. The display continually changes

   between 0000.01 and  0000.01 (not possible in PRESET-function). As soon as the resolution
   with the required counting direction appears release both keys.

Use with OPTO-RS 232 -plug:

The adapter plug opto-RS 232 permits the remote control of all key functions of the instrument from a
computer (PC or SQS-system).

11

Summary of Contents for 25 ES

Page 1: ...hl 6 Bedienung 7 Messbeispiele 7 Abbildungen 24 Table of Contents Page Designation 1 Description 9 Accessories 9 Introduction 9 Important notes concerning first time use 10 Technicaldata 10 Setting an...

Page 2: ...lbaren Messarme 844 Te Tx oder Aufnahmest cke 844 Tm umdrehbar Zubeh r Messarme 844 Te Erforderlich bei folgenden Messeins tzen 844 Tp 844 Ts 844 Tk 844 Tr 844 Tg Messarme 844 Tx Erforderlich bei folg...

Page 3: ...bers 25 ES empfehlen wir Ihnen diese Bedienungsanleitungaufmerksamundvollst ndigzulesen SicherheitshinweisezurBatterie nichtwiederaufladbar nicht ins Feuer werfen bittevorschriftsgem entsorgen Dieses...

Page 4: ...der Z hlrichtung Taste 2 3 und 2 4 gleichzeitig l nger als 2 Sek gedr ckt halten Die Anzeige wechselt st ndig zwischen 0000 01 und 0000 01 in der PRESET Eingabefunktion nicht m glich Sobald der Ziffer...

Page 5: ...n m glich 844 Tp 844 Ts 844 Tk 844 Tr nur Flanken 844 Tc 901 bis 908 911 H 912 913 2 Vorbereitung f r die Messung des Istma es mittels Einstellnormal z B Einstellehre 25 Eel Parallelendma oderEinstell...

Page 6: ...ription 1 Beam with measuring system 2 Right measuring arm holder adjustable with digital display and operating elements 2 1Housing 2 2LCD display 2 3 Key for ON OFF and HOLD 2 4 Selector key for meas...

Page 7: ...function press key 2 4 for more than 2 sec 12 Note The instrument cannot be switched off while the PRESET input function is active MIN MAX function Automatic storage and display of MIN and or MAX as...

Page 8: ...Tz 3 Measuringofdeviations comparativemeasurements 3 1 Mount suited measuring arms with measuring anvils in one of the mounting grooves of both measuring arm holders 3 2 Set display to nominal value u...

Page 9: ...errage pour maintenir le guide dans la position voulue 3 Guide gauche mobile 3 1 Vis de serrage pour maintenir le guide dans la position voulue permet pour chaque t che de mesure un quilibrage optimal...

Page 10: ...marche appuyer bri vement sur la touche 2 3 ON OFF Nota La derni re fonction activ e avant la mise hors circuit de l appareil appara t sur l affichage Mise hors circuit appuyer plus de 2 sec sur la t...

Page 11: ...ches 2 3 et 2 4 L affichage change constamment entre 0000 01 et 0000 01 impossible dans la fonction d introduction PRESET D s que la r solution est affich e avec le signe souhait rel cher les deux tou...

Page 12: ...de mesure en position dans l une des rainures des deux guides 3 2 Ajuster la cote th orique au moyen de l talon de r glage ou du normal proc der comme indiqu sous 2 2 2 4 3 3 Remettre z ro l affichage...

Page 13: ...F G H A B C D E TOL SET TOL MM SET INCH SET INCH SET REF I MM SET PRESET REF I SET MIN REF I HOLD B MM SET I Abbildungen Bitteausklappen Illustrations Please fold out Illustrations Ouvrir le volet 24...

Page 14: ...Notizen Notes Notes 23...

Reviews: