7
Mahr GmbH, MarSurf PS 10
Calibrating the probe
1. Using your finger nail, lift and remove the
standard
(11)
on the bottom of the MarSurf
PS 10
(see figure 1)
.
2. Read off the
Rz
nominal value from the
standard.
3. Place the MarSurf PS 10 on its vee-block
base so that the stylus tip is vertical.
4. Position the standard in such a way that the
entire traversing length is within the field
with the defined roughness
(see figure 3)
.
The traversing length will be
calculated automatically in accord-
ance with DIN EN ISO 4288 or
ASME
B46.1 depending on the
nominal value (see "MarSurf. Sur-
face Parameters" fold-out card pro-
vided). Half a pre-travel and post-
travel is added to this.
5. Set the height adjuster
(7)
on the MarSurf
PS 10 to the required height:
−
for the PHT 6-350 standard probe to ,
−
for other probes as required.
The probe must be parallel to the standard.
6. In the basic view tap the
button
(or
Calibrate
in the menu).
7. Select the calibration method.
Always select "Calibrate Rz" when using the
standard provided.
8. Enter the nominal value for the standard in
the "Nominal value" field.
9. Tap the
Start calibration
button.
Three measurements will be performed
auto matically.
10. After the third measurement a correction
value will be calculated from the three
meas ured values (
Rz
,
Ra
or
RSm
).
i
3
Getting started
You can start measuring as soon as you have
completed the following steps:
−
Remove the shipping clamp
−
Insert the drive unit into the MarSurf PS 10
(see
in Section 4)
−
Set the measuring conditions
−
Calibrate the probe
Removing the shipping clamp
You must remove the red shipping clamp
before using the MarSurf PS 10 (see in-
formation sheet 3763316)!
Make sure you keep the shipping clamp
in a safe place!
In the event that you need to return the
instrument, the shipping clamp must be
put into place again.
Switching on the MarSurf PS 10
1. Press the green
Start
key.
Performing the basic settings
1. In the "Congratulations" dialog:
•
Tap the
>
button to set the language
(
), followed by the date and time.
•
Select the unit of measurement "mm"
or "inch".
•
Tap
OK
.
The menu appears. Tap
OK
again to
display the basic view.
Setting the measuring conditions
1. In the basic view tap
(or the
Measuring
conditions
option in the menu).
2. Tap again and set the relevant meas ur-
ing conditions.
For details about the measuring conditions,
please see Section 5.
3. Tap
OK
. Tap
OK
again.
i
Summary of Contents for MarSurf PS 10
Page 6: ...Mahr GmbH MarSurf PS 10...
Page 185: ...3 Mahr GmbH MarSurf PS 10 1 2 3 4 5 6 VDE 7 8 Mahr GmbH i...
Page 251: ...3 Mahr GmbH MarSurf PS 10 1 2 3 4 5 6 VDE 7 8 Mahr GmbH i...
Page 273: ...3 Mahr GmbH MarSurf PS 10 1 2 3 4 5 6 VDE 7 8 Mahr GmbH i...
Page 295: ...3 Mahr GmbH MarSurf PS 10 1 2 3 4 5 6 VDE 7 8 Mahr GmbH i...
Page 381: ...Mahr GmbH MarSurf PS 10...
Page 382: ...Mahr GmbH MarSurf PS 10...
Page 383: ...27a 25 25 26 28a 27 28 29 30 Abb 6 Fig 6 Obr 6 Rys 6 6 6 bra ekil 6 6 6 6 Bild 6 Afb 6...
Page 384: ...www mahr com...