40
Mahr GmbH • Millimar C1200 M/T
ü
Ó
u
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
-0
1.000
Measurement
Display
Setup
System
Customer calibration
+ 0 0
1
. 0 0 0
3
Ó
ü
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
-0
1.000
Measurement
Display
Setup
System
Customer calibration
Calibration
Min
and accept
3
Ó
ü
q
p
Calibration Min
Nominal value
Calibration Max
-1.000
-0
1.000
Measurement
Display
Setup
System
Customer calibration
Calibration
Max
and accept
Measurement
Display
Setup
System
Absolute mode
Customer calibration
Factory calibration
Factory settings
Keyboard-/ menu interlock
Info
Ó
ü
q
p
ü
–
Enter value for
Calibration Max
(e.g. +1.000 mm). Press
u
to
select the required digit and press
q
p
to enter the required value.
–
Press
ü
to confirm
or
Ó
to cancel
the selection
–
Move the
Probe
to the
Min
position (e.g. measuring plate)
=>
The following appears on the
display:
Set probe to
Calibration Min
and
accept
–
Press
ü
to
confirm
–
Move the probe to the
Max
position (e.g. 2.000 mm gauge
block)
=>
The following appears on the
display:
Set probe to
Calibration Max
and
accept
–
Press
ü
to
confirm
=>
Probe calibration
according to
the customer's specification is
complete and activated.
–
Press the ON / OFF key to exit the
input menu.