43
Mahr GmbH • Millimar C1200 M/T
3
6
u
q
p
Calibration Min
Nennmaß
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
+
0 0 0 . 3 0 0
3
6
3
q
p
Calibration Min
Nennmaß
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
Please set the probe at
Calibration
Min
and accept
3
6
u
q
p
Calibration Min
Nennmaß
Calibration Max
-1.000
+0
1.000
Measurement
Display
Setup
System
Customer calibration
-
0 0 0 . 3 0 0
–
Enter value for
Calibration Min
(e.g. -300 µm). Press
u
to select
the required digit and press
q
p
to enter the required value.
–
Press
ü
to confirm or
Ó
to cancel
the selection
–
Enter value for
Calibration Max
(e.g. +300 µm). Press
u
to select
the required digit and press
q
p
to enter the required value.
–
Press
ü
to confirm or
Ó
to cancel
the selection
–
Connect the jumpers to the
Factory standard
as illustrated
The following appears on the
display:
Set probe to
Calibration Min
and
accept
–
Press
ü
to confirm
–
Connect the jumpers to the
Factory standard
as illustrated
For TESA probes, the minimum and maximum adjustable value for calibration
is 300 µm. The TESA adapter cable 7026704 must be connected between the
Millimar C1200 and the TESA probe. The display shows the value 784.4 mm