10
Mahr GmbH • Multimar 844 S
Lo
Li
10
A
20
20
10
8. Setting dimensional references
8.1 Setup for a Multimar 844 T
Required accessories:
-
844 Sp parallel jaw
10
in T-shape
-
Gage block combination
A
(in accordance with the reference dimension)
-
844 Sph Spacers
11
(for long jaws starting at 70 mm)
-
844 Sf Base plate
12
for vertical use (if necessary)
Handle the 844 Sf base plate carefully due to its heavy weight (7.5 kg)!
We recommend using the included table clamps to attach the base plate to the
measuring station.
Setting up a dimensional reference Lo for outer measurement:
Dimensional reference Lo = gage block combination (Li) + 844 Sp (2 x 20 mm)
Setting up a dimensional reference Li for inner measurement:
Dimensional reference Li = gage block combination between T-parallel jaws