Operation Manual MA 1  

 

8100216 Rev. 5 

13 

09/03/2023 

 

5 Operating the Device 

This section offers you information about:  

 

how to get started with the MA 1  

 

preparing the patient for testing  

 

performing a hearing  screening  

5.1 Getting Started with the MA 1 

5.1.1 Use of Equipment After Transport and Storage 

Make sure the device is functioning correctly before use. If the device has been stored in a 
colder environment (even for shorter time) allow the device to become acclimatized. This 
can take a long time depending on the conditions (like environmental humidity). You can 
reduce the condensation by storing the device in its original packaging. If the device is stored 
under warmer conditions than the use conditions no special precaution are required before 
use. Always ensure proper operation of the device by following routine check procedures 
for audiometric equipment. 

5.1.2 Switching the Device On and Off 

Switch on the device by pressing the black right/left button to turn on the MA 1.   

An automatic power shutoff occurs after 90 s of inactivity to conserve battery life.   

5.2 Front Panel Controls  

                                

 

Figure 3 

 
Table 4 Explanation of Front Panel 

No. 

Function 

 

 

 

 

Power button and R/L Ear Selector 

LED Ear Indicator, R = right, L = Left 

LED Level Indicator 

dB Level selector buttons 

Frequency selector buttons/stimulus presentation buttons  

Summary of Contents for MA 1

Page 1: ...Operation Manual MA 1...

Page 2: ......

Page 3: ...ng and Disinfection Recommendations 8 3 4 Disposables 9 3 5 Components Replacement Parts 9 3 6 Recycling and Disposal 9 4 Unpacking and Installation 10 4 1 Unpacking the System 10 4 2 Hardware Orienta...

Page 4: ...www maico diagnostics com german support resources International https www maico diagnostics com support resources Copyright 2023 MAICO Diagnostics All rights reserved No part of this publication may...

Page 5: ...ENTs hearing healthcare professionals or other trained technicians in a hospital clinic or healthcare facility 1 3 Contraindications of Use The patient is too young sick or uncooperative to perform th...

Page 6: ...ystem including safety information as well as maintenance and cleaning recommendations READ THIS ENTIRE OPERATION MANUAL BEFORE ATTEMPTING TO USE THIS SYSTEM Use this device only as described in this...

Page 7: ...is manual accompanying labels and or inserts A defective product should not be used Make sure all connections to external accessories are snug and secured properly Parts which may be broken or missing...

Page 8: ...tion consult accompanying documents Warning consult accompanying documents Return to authorized representative special disposal required Reference number Medical Device Global Trade Item Number Patien...

Page 9: ...calibration Do not use the device if any damage is suspected Calibration of the device The device and the transducers complement each other and share the same serial number i e MA7663252 Therefore the...

Page 10: ...device adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this device and the other equipment should be observed to v...

Page 11: ...t of repair during the guarantee period please enclose evidence of purchase with the device 3 2 Maintenance In order to ensure that the device works properly it has to be checked and calibrated at lea...

Page 12: ...lies that are supplied with your device Ear cushion covers are intended for single use only These should be discarded after use They cannot be cleaned In case of re use of the single use disposables y...

Page 13: ...onent appears to be damaged in shipment contact your distributor immediately to report it Do not attempt to use any component or device that appears to be damaged Reporting Imperfections Notify the ca...

Page 14: ...iometric Headset 2x AA Batteries Carrying Case Operation Manual Quick Guide Audiogram Pad Applied parts according to IEC 60601 1 As download from the download center see accompanying leaflet Table 3 R...

Page 15: ...attery life The device will power off after 90 s of inactivity When batteries get too low to produce the proper signal remaining battery power of 2 1 V or lower the device will shut off and will not t...

Page 16: ...condensation by storing the device in its original packaging If the device is stored under warmer conditions than the use conditions no special precaution are required before use Always ensure proper...

Page 17: ...r Conduction Testing 5 4 1 Screening A hearing screening utilizes a Pass or Refer result and is used to determine if further testing is required as a hearing problem may exist Patients are typically s...

Page 18: ...ither ear would result in a Refer Note When presenting tones be sure to vary the timing of the tone presentation This is to prevent the patient from routinely raising his her hand instead of respondin...

Page 19: ...e disposal of authorized service companies STANDARDS Safety Standards IEC 60601 1 2005 Cor 2006 Cor 2007 A1 2012 Internally powered Type B Applied Parts EMC Standard IEC 60601 1 2 2014 The MA 1 is not...

Page 20: ...ansducers Headband tension DD45 Headband Static Force 4 5 N 0 5 N DD65 v2 Headband Static Force 10 0 N 0 7 N Inputs Tone Accuracy Level 4 dB Precision 5 dB level steps Outputs Left Right Stimuli Prese...

Page 21: ...al MA 1 8100216 Rev 5 18 09 03 2023 6 2 Connections Figure 5 CONNECTIONS No Connection socket Specification 1 Phone ZA 10 UA 7 Veff 6 3 Pin Assignment SOCKET CONNECTOR PIN 1 PIN 2 Phone 3 5 mm Mono Gr...

Page 22: ...lent threshold sound pressure level RETSPL dB re 20 Pa DD45 DD65 v2 500 7 0 26 1 1000 15 0 32 4 2000 26 0 43 6 4000 32 0 43 8 Table 7 Reference Values for Stimulus Calibration REFERENCE VALUES FOR STI...

Page 23: ...o cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The MA 1 is suitable for use in all commercial industrial business and residential environments Harmonic emissions...

Page 24: ...the electromagnetic environment specified below The customer or the user of the MA 1 should assure that it is used in such an environment Immunity test IEC EN 60601 test level Compliance level Electr...

Page 25: ...equirements for Basic Safety and Essential Performance Collateral Standard Electromagnetic Compatibility Requirements and tests 3 DIN EN ISO 14971 2012 Application of risk management to medical device...

Page 26: ...Operation Manual MA 1 8100216 Rev 5 23 09 03 2023 6 7 Checklist for Subjective Device Check...

Page 27: ...ecifications are subject to change without notice MAICO Diagnostics GmbH Sickingenstr 70 71 10553 Berlin Germany Tel 49 30 70 71 46 50 Fax 49 30 70 71 46 99 E mail sales maico biz Internet www maico b...

Reviews: