Instruction Manual 

PowerPoint Power Supply,   

Model MP-300 

   

 
 
 
 

 
 

 
 

 

 

 
 
 
 

Summary of Contents for MP-300

Page 1: ...Instruction Manual PowerPoint Power Supply Model MP 300...

Page 2: ...ents Packing list 3 Warning 4 Safety Information 5 Product Specifications 7 Product Description 8 Controls and Features Installation Instructions 10 11 Operation Instructions 11 Troubleshooting Guide...

Page 3: ...3 Packing list MP 250 or CS 300 1 x MP 300 1 x Power Cord for power supply 1 x Extra fuse inside power plug on the rear of unit 1 x power supply instruction manual...

Page 4: ...Do not alter the power supply series Failure to adhere to these directions could result in personal and or laboratory hazards as well as invalidate the equipment warranty 3 Use a properly grounded ele...

Page 5: ...elow to ensure safe operation of the unit The MP power supply series has been designed for use with shielded wires thus minimizing any potential shock hazard to the user We recommend against the use o...

Page 6: ...let the unit equilibrate to room temperature for a minimum of 2 hours before using it 6 Before using any cleaning or decontamination method except those recommended by the manufacturer users should ch...

Page 7: ...r 1 999 mins with alarm continuous Safety Device No Load detect Leakage detect Overload detect Over temperature protection Shrouded plugs and sockets Crossover Yes Operating Temperature Ambient 400 C...

Page 8: ...d operation of the 300V Power Supply including important information on safety and maintenance of the unit The MP 300 power supply are capable of running DNA RNA electrophoresis Electroblotting SDS PA...

Page 9: ...NA and protein electrophoresis Better design for refrigerated condition operation Introduction Overview MP 300 Power Supplies are recognized as one of the most advanced power supply equipped with outs...

Page 10: ...either Voltage or Time value 3 Key to activate or stop the unit 4 Key to temporarily interrupt power to an operation in progress without terminating electrophoresis and to resume power after pausing w...

Page 11: ...e back of the unit 3 Turn the AC power ON by the switch on the rear of the machine 4 Connect the DC output jacks from the electrophoresis unit insert the red lead into the red output jack and the blac...

Page 12: ...o stop when the timer is up 8 Press Key to start the run The LED is lit 9 Press Key to temporarily interrupt the run if necessary The LED is lit 10 Press Key again to stop the unit at any time if nece...

Page 13: ...ce the fuse Repeated fuse broken Hardware failure Contact your suppliers service department Operation stops with alarm The screen displays Electrophoresis leads are not connected to the power supply o...

Page 14: ...power on off and then restart application If the power supply cannot be restarted disconnect the power supply from the mains outlet and contact Technical Service Err2 power supply stops alarm sounds V...

Page 15: ...ower supply from the mains outlet and contact Technical Service Err7 power supply stops alarm sounds Poor connection between power cable and power supply damaged power cable or electrode Check power c...

Page 16: ...ts where repair or modification has been performed by anyone other than your dealer or an appointed distributor or representative are no longer under warranty from the time the unit was modified Units...