137
7 Cleaning & Maintenance
The painted surfaces of the filter areas on the built-in UV Transilluminator
must be cleaned with water and soap, using a sponge or towel. Dry the filter
surface with a soft cloth after each operation. Never use abrasive cleaners,
solvent based cleaners or scouring pads. The housing may be cleaned with a
moist cloth containing a mild soap solution.
Always disconnect the SmartView Pro Imager System from the electrical
power prior to cleaning.
7.1 Replacing the Fuse
For additional fuses, contact MS local distributer.
To replace the fuse:
1. Turn off the main power switch at the rear of system and detach the power
cord.
2. Open the fuse compartment located inside the Power Entry Module by
inserting a small flathead screwdriver into the slot below the ON/OFF
switch. Turn the screwdriver to gently pry open the fuse compartment.
Note: the fuse compartment will not open with the power cord in place.
3. Pull the fuse holder out of the compartment and inspect the fuse. If the fuse
is burned or there is a break in the fuse element, replace the fuse with an
identical type of fuse (
T2A/250V~
) as provided in the fuse holder (see
figure below)
4. Place the fuse holder back into the compartment.
5. Snap the cover close.