background image

                                                

 

 

Note: 

 

Make sure the power is cut off and all capacitors need to be discharged under 

 

this measurement. 

 
3.6    Audible continuity Test 

1)    Set the selector switch to desired “ ” position. 
2)    Connect the test leads to two point of circuit, if the resistance is lower than 
 

approx. 50Ω, the buzzer sounds.

 

 

Note:   

 

Make sure the power is cut off and all capacitors need to be discharged under 

 

this measurement. 

 
3.7    Battery Test 

1)    Set the selector switch to desired “

1.5V

” or “

9V

” position. 

2)    Connect the probe across the source or load under measurement. 
3)    Read the result from the LCD panel. 
 

4.    Battery replacement 
 

1)    When the battery voltage drop below proper operation range, the “ ” symbol 

 

will appear on the LCD display and the battery need to changed. 

2)    Before changing the battery, set the selector switch to “

OFF

” position to power 

 

off. 

3)    Remove the screw on the battery cover and lift the battery cover. 
4)    Replace the old battery with the same type battery. 
5)    Close the battery cover and fasten the screws. 

 

Caution:   

 

Dispose the used batteries according to the rules, which are defined by each 

 

community. 

 

Warning:

   

 

If an explosion or fire hazard could occur through fitting a battery of the wrong 

 

type. 

 

5. 

Fuse replacement 

 

1)    This meter is provided with a F0.5A/600V fuse to protect the resistance, Diode,

 

 

Audible continuity, battery test and the current measuring circuits. 

2)    Ensure the meter is not connected to any external circuit, set the selector switch 

 

to “

OFF

” position to power off. 

3)    Remove the screw on the battery cover to lift the battery cover, and remove the 

 

two screws on the bottom case, loose the button stand upward side the battery 

Summary of Contents for MTD30

Page 1: ...ection for electrical and electronic equipment The RESPONSIBLE BODY shall be made aware that if the instrument is used in a manner not specified by the manufacturer the protection provided by the instrument may be impaired The finger or any part of your body shall not be beyond the barrier of the test probe when measuring Individual protective equipment must be used if HAZARDOUS LIVE parts in the ...

Page 2: ... data 1 8 Use the DMM indoor altitude up to 2000m and temperature 5 to 40 Maximum relative humidity 80 for temperatures up to 31 decreasing linearly to 50 relative humidity at 40 Pollution Degree 2 2 SPECIFICATIONS 2 1 GENERAL SPECIFICATIONS Display LCD with a max reading of 1999 Range control Manual range control Polarity Automatic negative polarity indication Zero adjustment Automatic Over range...

Page 3: ...2 2 AC Voltage Range Accuracy Resolution 200V 1 5 3 100mV 500V 1V Average sensing calibrated to rms of sine wave Frequency 40 500Hz Overload protection 500V DC or AC rms Impedance 450kΩ 2 2 3 DC Current Range Accuracy Resolution 2000μA 1 5 2 1μA 20mA 10μA 200mA 2 0 2 100μA Overload protection F0 5A 600V fuse 2 2 4 Resistance Range Accuracy Resolution 200Ω 1 0 5 0 1Ω 2000Ω 1 0 3 1Ω 20kΩ 10Ω 200kΩ 1...

Page 4: ...here the value of voltage is needed 3 Read the result from the LCD panel The polarity of the red lead connection will be indicated along with the DC voltage value Note a If the voltage range is not known beforehand set the selector switch to high range and work down b When 1 or 1 is display over range is being indicated and the selector switch must be set to a higher range c Don t apply more than ...

Page 5: ...circuits 3 4 Resistance Measurement 1 Set the selector switch to desired Ω position 2 Connect tip of the test leads to the points where the value of the resistance is needed 3 Read the result from the LCD panel Note a If the resistance value being measured exceeds the maximum value of the range selected an over range indication will be displayed 1 select a higher range For resistance of approximat...

Page 6: ...ttery need to changed 2 Before changing the battery set the selector switch to OFF position to power off 3 Remove the screw on the battery cover and lift the battery cover 4 Replace the old battery with the same type battery 5 Close the battery cover and fasten the screws Caution Dispose the used batteries according to the rules which are defined by each community Warning If an explosion or fire h...

Page 7: ... case is closed 2 To avoid contamination or static damage do not touch the circuit board without proper static protection 3 If the meter is not going to be used for a long time take out the battery and do not store the meter in high temperature or high humidity environment 4 Repairs or servicing not covered in this manual should only be conducted by a qualified person 5 Periodically wipe the case ...

Reviews: