2.2.4 Resistance 

Range 

Accuracy 

Resolution 

  200Ω 

± ( 1.0%+5 ) 

0.1Ω 

  2000Ω 

 

± ( 1.0%+3 ) 

 

      1Ω 

20kΩ 

      10Ω 

200kΩ 

100Ω 

2000kΩ 

± (1.5%+3 ) 

      1kΩ 

Overload protection: F0.5A/600V fuse 
 

2.2.5 Diode and Audible continuity test 

Range 

Description 

Test condition 

   

Display    read 
approximately 
forward voltage 
of diode 

Forward DC current 
approx. 10μA 
Reversed DC voltage 
approx. 1.8V 

 

Built-in buzzer 
sounds     if 
resistance  is 
less than 50Ω 

Open circuit voltage 
approx. 1.8V 

Overload protection: F0.5A/600V fuse 
 

3.    OPERATION 
 
3.1    DC Voltage Measurement 

1)    Connect the black test lead to the "

COM

" socket and red test lead to the 

 

"

VΩmA

" socket. 

2)    Set the selector switch to desired “

V

” position. 

3)    Measure the voltage by touch the test lead tips to the test circuit where the 

 

value of voltage is needed. 

4)    Read the result from the LCD panel. The polarity of the red lead connection will 

 

be indicated along with the DC voltage value. 

 

Note: 
 

a)     If the voltage range is not known beforehand, set the selector switch to high 

 

range and work down. 

b)    When “1” or “-1” is display, over-range is being indicated and the selector switch 

 

must be set to a higher range. 

Summary of Contents for MTD33

Page 1: ...tion for electrical and electronic equipment The RESPONSIBLE BODY shall be made aware that if the instrument is used in a manner not specified by the manufacturer the protection provided by the instru...

Page 2: ...titude up to 2000m and temperature 5 to 40 Maximum relative humidity 80 for temperatures up to 31 decreasing linearly to 50 relative humidity at 40 Pollution Degree 2 2 SPECIFICATIONS 2 1 GENERAL SPEC...

Page 3: ...tection 500V DC or AC rms Impedance 1M 2 2 2 AC Voltage Range Accuracy Resolution 200V 1 5 3 100mV 500V 1V Average sensing calibrated to rms of sine wave Frequency 40 500Hz Overload protection 500V DC...

Page 4: ...0 5A 600V fuse 3 OPERATION 3 1 DC Voltage Measurement 1 Connect the black test lead to the COM socket and red test lead to the V mA socket 2 Set the selector switch to desired V position 3 Measure the...

Page 5: ...3 Remove power from the circuit under test and open the normal circuit path where the measurement is to be taken Connect the meter in series with the circuit 4 Read the result from the LCD panel The...

Page 6: ...measurement display shows the approx forward voltage of this diode Note Make sure the power is cut off and all capacitors need to be discharged under this measurement 3 6 Audible continuity Test 1 Con...

Page 7: ...ws on the bottom case loose the button stand upward side the battery box of the front case and lift the bottom case 4 Replace the old fuse with the same type and rating 5 20mm F0 5A 600V or 5 20mm F10...

Reviews: