Here it is possible to change the parameters for the two different CX filters,
the FIR filter and the Time Gain filter. If one of the buttons are pressed then
the Trace/Radagram View will show how the choice of filter parameters
affects the trace otherwise the Trace/Radagram View will show the
trace/radargram only with DC removal filter.
If the FIR switch is OFF then DC Removal filter is ON for all displayed data.
If the FIR switch is ON this filter will be applied on all displayed data
according to the settings made. The FIR filter (Finite Impulse Response
Filter) is essentially a band-pass, time domain filter, removing unwanted
noise from the signal. the default parameters 25 and 5 will be appropriate.
Observe that the setting of the option BKG Removal only affects the FIR
filter and works on a trace level as a high-pass filter. It will not affect the
background removal option in the main menu for 2D Projects.
The time gain filter is applied to compensate for the geometrical and
conductive attenuation of the signal in matter. The intensity of the filter is
set in the main menu for 2D projects while the settings are changed here.
The maximum linear coefficient can be set to max 3000 and the maximum
exponential coefficient to max 1000.
In the
Wheels Menu
user defined wheels can be administrated; added,
edited or removed. Note! The predefined wheels can not be removed or
modified.
In
GPS settings
(only available if the GPS functionality is enabled) the
GPS settings are seen: