Addendum 0078-1.0
Page 3
Positioning the Instrument
Warning!
Do not position the instrument such that the power cord, where it exits the
product, is unreachable for disconnection.
Warning!
Do not obstruct the ventilation slots underneath the instrument, nor the
fans on the rear panel. Restricting airflow can damage the instrument or
cause overheating.
Replacing the system fuse
Warning!
Fuses must not be replaced by the operator. Only the supervisor or a
Malvern representative should attempt to change the fuse.
If the instrument does not power up, check the system fuses. These are located in
the mains power switch on the rear panel.
Before changing a fuse, disconnect the instrument from the mains power.
Pull the fuse holders out and replace any faulty fuses with ones of the following
specification:
Positioning the Instrument
- Read in place of relevant section on page 3-1 of
the Zetasizer μV and Zetasizer APS Essentials Manual.
Replacing the system fuse
- Read in place of relevant section on page 4-3 of
the Zetasizer μV and on page 4-2 of the Zetasizer APS Essentials Manual.
Rating
:
Size
:
T 2A HRC 250v (T = Time delay)
5mm x 20mm