General cells and cuvettes
Chapter 2
Zetasizer Nano accessories guide
Page 2-7
Use, cleanliness and filling advice
Note
Before filling and using a cell or cuvette, consult the
cleaning
section for
each cell or cuvette, and perform any cleaning and maintenance proce-
dures described.
When filling the cell there are several actions to consider; some that apply to all
cells and others that are only applicable to the measurement type and the cell
chosen.
The cleanliness of the cells used in each measurement is a paramount importance.
As well as the information below, refer to the respective
cleaning cells/cuvettes
section
for each cell, where more specific advice will be given.
Size cells and cuvettes
All
size
cells should be rinsed/cleaned with filtered dispersant before use.
Molecular weight cells and cuvettes
All
molecular weight
cells should be rinsed/cleaned with the filtered standard
(e.g. Toluene) or solvent, then dried in a dust free environment such as a lami-
nar flow cabinet, before use.
Zeta potential cells
All
zeta potential
cells should be rinsed/cleaned with filtered dispersant
before use.
Additionally
The cell should be filled slowly to avoid air bubbles from being created. Ultra-
sonication can be used to remove air bubbles - but only if the sample is suitable
for use with ultrasonics.
If using syringe filters for the dispersant, never use the first few drops from the
syringe, in case there are any residual dust particles in the filter that may con-
taminate the dispersant.
Summary of Contents for Zetasizer Nano Series
Page 2: ......
Page 3: ...Zetasizer Nano accessories guide MAN0487 Issue 1 1 April 2013...
Page 12: ...Chapter 1 Introduction and accessory range Page 1 6 MAN 0487...
Page 42: ...Chapter 2 General cells and cuvettes Page 2 30 MAN 0487...
Page 98: ...Chapter 5 Microrheology Page 5 20 MAN 0487...
Page 125: ......