General cells and cuvettes
Chapter 2
Zetasizer Nano accessories guide
Page 2-17
1.
Hold the cell near the top, away from the lower measurement area, and push
into the cell holder until it stops
.
2.
The cells will fit in the cell either way round but do have a correct orientation.
Please refer to the section below.
3.
Close the cell area lid
.
Note
Different versions of the capillary cells have unique thermal plates. If
required, please contact your Malvern representative for the correct plates
for your cell.
Folded capillary cell - orientation and insertion
The clear folded capillary zeta potential cells can show significant differences in
sample count rate depending on the orientation of the cell in the cell holder. In
most cases the difference does not affect the quality of the result, only the
attenuator selection. However, in extreme cases, where the sample being measured
is a poor scatterer, the measurement may not be possible in one of the orientations.
The diagram below shows the cell with the
preferred
orientation in the cell
holder.
DTS1070 DTS1060/61
ill 8798 / 7945
DTS1070 cell
When inserting the cell, ensure that the
Malvern logo
faces towards the front of
the instrument. Press down until the cell clicks into place.