Chapter 2
General cells and cuvettes
Page 2-22
MAN 0487
ill 8508
Note
With the procedure complete, the measurement face of the cuvette (some
have a small
triangle
at the top of the cell) and the coloured band on the
Dip cell label must face in the same direction. This is to ensure the orienta-
tion is correct when inserted into the cell holder.
Removing the Dip cell
With care, by simultaneously holding the base of the Dip cell cap and the top of the
cuvette, both the Dip cell and cuvette can be removed together. If adequate
purchase cannot be obtained on both parts, then the following procedure is
recommended.
Lift the Dip cell up out of the cuvette, but before completely removing, gently
tap the bottom of the Dip cell on the top of the cuvette
. This will dislodge
any remaining drops of sample from the cell into the cuvette.
If the Dip cell is simply lifted out of the cuvette there is a risk of drops of sam-
ple falling from the bottom of the Dip cell onto the instrument and surround-
ing area. This is especially important when using solvent based samples.
Stop
Stop
max
45°
10mm
max
Front
Front
Front
1
2
3
4
5
6
Coloured
band
Triangle