Microrheology
Chapter 5
Zetasizer Nano accessories guide
Page 5-9
For a description of the other measurement parameters, refer to the
Measurement
- Advanced
description in
the size SOPs
section of the main user manual.
For all other microrheology SOP windows, refer to the size SOP
section in the main user manual.
Microrheology measurement process
With the cell loaded into the instrument, and the SOP configured, a measurement
can be performed.
Overall to make a microrheology measurement, only one measurement is
technically needed - this is a size measurement that the software uses to calculate
the relaxation times, and therefore the viscosity and moduli values.
However, in order to get to this point, and in order to make reliable measurements,
the full microrheology measurement process consists of several basic steps, these
are described in the following paragraphs below. As the measurement progresses
through the measurement steps, the software will prompt instructions for the user
to perform where relevant.
Each instruction will involve a manual operation and the user must open the cell
area lid, perform the specified action, then close the lid again before continuing.
The process a microrheology measurement performs is shown in the optimization
stage of the SOP.