Chapter 4
Maintenance
Page 4-4
MAN 0382
Cleaning the High concentration cell
General cleaning
Rinsing of the cell
prior
to a measurement should be carried out by flushing
through with copious amounts of de-ionised water.
External surfaces of the assembled cell can be wiped clean with a weak soap solu-
tion
For more intensive cleaning follow the instructions below.
Intensive cleaning
The cell first has to be disassembled before cleaning can be performed.
ill 8447
Remove the screw cap
.
Separate the two halves of the cell
by pulling the rear casing vertically away
from the metal front.
Note how the Electrode chambers and quartz measurement cell block are
assembled
.
Remove the chambers and cell block from the metal front casing
.
Detach the pipework and remove the top port
.
Protect the cell block from damage
.
1
2
3
4
5
6