Peristaltic pump is under pressure! Do not open housing!
A peristaltic pump might be part of a system, which is under pressure.
Check pressure before opening peristaltic pump, and adjust pressure to
atmosheric pressure.
Flexible tube, conveying belt, contact pulleys and contact springs are the only parts of the pump subject
to wear. They are simple to change.
Note
If you send back the peristaltic pump to the M&C service for repair, please
let us know what kind of condensate has been pumped.
Before sending the pump back clean all parts from dangerous or highly
aggressive contaminants.
15.2.1 CHANGING THE PUMP TUBING
1
Conveying belt
2
S-bolt
3
Tubing set
4
contact pulley
5
springs
Figure 8
Changing the pump tubing
For changing the pump tubing please proceed as follows:
1. Unplug the pump from the mains voltage. The device needs to be voltage free.
2.
Open tube connections at the pump;
3. Press conveying belt
at the recessed grips and turn S-bolt
clockwise up to limit stop;
4. Take away conveying belt
and remove the old tubing set
from the guides by pulling on the
tube connectors;
5. Press the two contact pulleys
and check whether the spring pressure is still sufficient, if not,
the contact springs have to be changed (see chapter 15.2.2);
6. Put the new tubing set
with the tube connectors into the guides of the conveying belt
;
Note
Only the usage of the original tubing set guarantees a proper functionality.
Never lubricate the tube.
Before mounting the pump check all parts for contaminations and clean if
necessary.
Summary of Contents for 03G6000
Page 44: ...44 SS M05_Marine 1 01 00 www mc techgroup com Figure 13 Wiring diagram part 1 ...
Page 45: ...www mc techgroup com SS M05_Marine 1 01 00 45 Figure 14 Wiring diagram part 2 ...
Page 46: ...46 SS M05_Marine 1 01 00 www mc techgroup com ...
Page 47: ...www mc techgroup com SS M05_Marine 1 01 00 47 ...
Page 48: ...48 SS M05_Marine 1 01 00 www mc techgroup com ...