40
SS-M05_Marine | 1.01.00
www.mc-techgroup.com
Indication
Problem
Possible Cause
Check/Solution
Clean clogged line or replace;
Gas flows?
Alarm LED on
the LA electro-
nics is red
Optional liquid alarm
sensor:
Sensor turns off
sample gas pump
automatically
Cooler is overloaded momentarily due to
excessive amount of condensate;
OK?
WARNING! Aggressive condensate! Use
personal protective equipment!
Check tubing
for condensate removal;
OK?
Check peristaltic pump tubing (see instruction
manual peristaltic pump SR25.1);
OK?
Check peristaltic pump SR25.1 (see manual
peristaltic pump SR25.1);
OK?
Optional flowmeter:
Needle valve closed
Check cooler according to instruction manual;
Adjust needle valve to the desired flow
Middle LED
on cooler is
green
Cooler and
sample gas
pump are
working;
condensate
in sample
gas line
Peristaltic pump
tubing defective
Peristaltic pump not working?
WARNING! Aggressive condensate! Use
personal protective equipment!
Change
pump tubing (see instruction manual peristaltic
pump SR25.1);
OK?
Peristaltic pump
SR25.1 defective
Check peristaltic pump (see instruction manual
peristaltic pump SR25.1)
OK?
Sample gas is not
sufficiently dry
Check cooler (see instruction manual
ECP1150M);
LED of the LA
electronics is
green
Optional liquid alarm
sensor:
Sensor did not turn
off pump.
Check the LA electronics and replace if
necessary
17
DISPOSAL
If the device has reached the end of its life cycle, please observe the legal requirements.
regulations and any other existing standard regulations in your country.
Summary of Contents for 03G6000
Page 44: ...44 SS M05_Marine 1 01 00 www mc techgroup com Figure 13 Wiring diagram part 1 ...
Page 45: ...www mc techgroup com SS M05_Marine 1 01 00 45 Figure 14 Wiring diagram part 2 ...
Page 46: ...46 SS M05_Marine 1 01 00 www mc techgroup com ...
Page 47: ...www mc techgroup com SS M05_Marine 1 01 00 47 ...
Page 48: ...48 SS M05_Marine 1 01 00 www mc techgroup com ...