Specifications
Airflo 24
Weight:
Output
pressure:
Batteries:
Battery Life:
4.1Kg
0.3
5
bar
1 x 24v - 2200mAh Ni-MH Airflo 24 Battery Pack (Mangar Prt No.
CD0314
)
Greater than 500 charging cycles (2
years depending on usage)
Environmental protection: IP41 (Protection from entry by solid objects with a diameter or
thickness greater
than 1.0mm & dripping water)
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislation
Charger:
Model number
- HX2902SR
Input:
Output:
AC 100v - 240v, 50-60Hz, 0.5A Max
DC
30v, 0.45A
Class II Double insulated equipment
.
Trouble shooting guide
Problem
Remedy
•
Compressor does not run when the 4 way
hand
control buttons are pressed.
•
Ensure that the Airflo
24 has been
switched
on.
•
Ensure the charger is disconnected from the
Airflo
•
Check that there are no kinks in the air hose.
•
Compressor keeps stopping when trying to
inflate the lifting aid.
•
Check that there are no kinks in the air hose.
•
Bottom Battery status indicator Flashing Red
every 2 seconds
•
Compressor causes electrical interference
with
another piece of electrical equipment.
•
Move either the compressor or other item
further away from each other until the
interference stops.
English
Airflo
24
•
Bottom Battery status indicator solid Red
Light
•
Low Battery Capacity, recharge.
•
Bottom Battery status indicator flashing Red
every
½ sec
ond
•
Internal Pump Temperature to
o
high – allow
Airflo
to cool down 60-90mins before
attempting
a lift.
•
No Battery status indicator Lights
•
Unit switched off -turn unit on.
•
Thermal fuse on battery has tripped – Allow
unit
to cool down 60-90min
utes
s. A solid red
LED will
display after temperature
normalizes and fuse
resets. This will have
reset the charge sequence.
To avoid damage to the Airflo after 12
hour
charge,
disconnect from charger and allow
another 60-90
minutes
for battery cool
down.
•
Battery Voltage low – Re charge. If occurs
repeatedly replace the battery.