Healthcare solutions designed for life
5
Quick Start Guide
DURING A LIFT
AFTER COMPLETING A LIFT
Disconnect the inflatable unit and hand control and pack away. Turn off the compressor
and
the bottom green light will go out.
The product will also auto turn off after 2.5 minutes
of inactivity.
Press the Battery Status Indicator Button.
4 GREEN
OK to Use
GREEN
GREEN
GREEN
GREEN
3 GREEN
OK to Use
GREEN
GREEN
GREEN
2 GREEN
OK to Use
GREEN
GREEN
1 GREEN
OK to Use
GREEN
RED
DO NOT USE
RECHARGE
RED
Each time the hand control is operated the unit does a battery test and indicates the
battery state. The hand control button is then released when a compartment in the lifting
device is filled.
4 GREEN
OK to Use
GREEN
GREEN
GREEN
GREEN
3 GREEN
OK to Use
GREEN
GREEN
GREEN
2 GREEN
OK to continue
the lift
GREEN
GREEN
1 GREEN
OK to continue
the lift
GREEN
RED
OK to continue
the lift - Recharge
when the lift is
complete
RED
If you see a single solid red light (if this is within a lift - the lift can be completed), the Airflo
should be put on unbroken charge immediately for a minimum of
3
hours. This will provide
sufficient capacity to complete one more lift.
Once the red light shows to achieve full capacity a full 12 hour charge is required.
English