9
A
B
A
3
•
Connect the four hoses (
A
) from the
4-Way Hand Control to the corresponding
numbered and coloured connectors on the
Mangar Camel or ELK.
•
Connect the hose (
B
) from the hand control
to the Airflo
24 air outlet socket (
C
). The
connector will ‘click’ into place when firmly
pushed onto the socket.
•
To switch on, press the ‘Power’ button (
D
).
The compressor will run when a button on
the hand control is pressed and stop when
the button is released. An internal pressure
switch will stop the compressor running
when each corresponding section of the
inflatable product has reached full pressure.
C
D
Mains Power Adaptor
(optional)
An optional mains power adaptor is available to
power the Airflo 24 directly from a mains power
socket. The adaptor is specific to the Airflo 24
and is the only approved unit to be used for
this purpose. Connect the adaptor between a
mains power socket and the adaptor inlet
socket on the end of the Airflo 24 (Fig. 2).
NOTE:
This adaptor, has a non-standard output
connector and powers the Airflo 24 directly,
it will not charge the battery. The only way to
charge the battery is to use the charger which
was supplied with the Airflo 24. With the power
adaptor connected, the Airflo 24 behaves
in the same way as if it was running on the
battery, the only functional difference is
that the battery indicator will
show bottom
LED
orange whilst the Airflo is running
.
With
the adaptor connected, the Airflo 24 will
also run if there is no battery fitted.
Operation
- fig.
Disconnect the Airflo
24 from the battery
charger and place it in a safe and convenient
position.
NOTE:
The Airflo compressor should
not be placed in a position where it is difficult
to operate the ‘Power’ button in an emergency.
2
•
Pressing any of the numbered hand control buttons will deliver air to the corresponding
numbered hose (follow the User Instructions for the Mangar Health product that is to be
inflated).
•
To switch the Airflo 24 off after use, press the ‘Power’ button. The bottom indicator light will go
out.
The pump will also automatically tu
r
n off after 2.5 minutes of inactivity.
English
Airflo
24
3