Press the hand control or 

Airflo 12 built-in buttons 

to raise or lower the 

lifting equipment.

3

Connect the  hose 

from the  Mangar 

pneumatic lifting 

equipment.

2

Connect the Hand 

Control.

1

Operating

WARNING:

 

The battery must be put on charge immediately as soon as the red indicator light 

appears. Running the Airflo 12 on RED until the battery is exhausted will damage the battery.

TO DETERMINE THE BATTERY CHARGE STATE, YOU MUST DISCONNECT THE CHARGING LEAD 

AND THEN PRESS THE ‘BATTERY STATUS INDICATOR’ BUTTON TO ACTIVATE THE BATTERY 

STATUS INDICATOR LIGHTS:

Battery Status 

Indicator 

Button

RED

Recharge

Immediately

2 GREEN

Recharge 

Soon

3 GREEN

OK to Use

4 GREEN

OK to Use

1 GREEN

Recharge 

Before Next Use

IMPORTANT: 

The Airflo 12 is designed not to deflate if it’s internal processor can see that there 

would be insufficient battery voltage to fully inflate the lifting product. This safety feature stops 

an inflatable bathing product from being lowered to the bottom of a bath if there is insufficient 

battery voltage to raise it again. The safety feature means that once you have inflated the 

product using the ‘Up’ button, you must wait 30 seconds before pressing the ‘Down’ button. 

If the ‘Down’ button is pressed too soon, you may see the battery status indicator track down 

quicker to a RED light and Deflate Lock. Unlock by putting the Airflo 12 back on charge for 2 

hours.