2
Press the hand control or
Airflo 12 built-in buttons
to raise or lower the
lifting equipment.
3
Connect the hose
from the Mangar
pneumatic lifting
equipment.
2
Connect the Hand
Control.
1
Operating
WARNING:
The battery must be put on charge immediately as soon as the red indicator light
appears. Running the Airflo 12 on RED until the battery is exhausted will damage the battery.
TO DETERMINE THE BATTERY CHARGE STATE, YOU MUST DISCONNECT THE CHARGING LEAD
AND THEN PRESS THE ‘BATTERY STATUS INDICATOR’ BUTTON TO ACTIVATE THE BATTERY
STATUS INDICATOR LIGHTS:
Battery Status
Indicator
Button
RED
Recharge
Immediately
2 GREEN
Recharge
Soon
3 GREEN
OK to Use
4 GREEN
OK to Use
1 GREEN
Recharge
Before Next Use
IMPORTANT:
The Airflo 12 is designed not to deflate if it’s internal processor can see that there
would be insufficient battery voltage to fully inflate the lifting product. This safety feature stops
an inflatable bathing product from being lowered to the bottom of a bath if there is insufficient
battery voltage to raise it again. The safety feature means that once you have inflated the
product using the ‘Up’ button, you must wait 30 seconds before pressing the ‘Down’ button.
If the ‘Down’ button is pressed too soon, you may see the battery status indicator track down
quicker to a RED light and Deflate Lock. Unlock by putting the Airflo 12 back on charge for 2
hours.