Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
40
Operation in the carry bag
Operation
5.5
Operation in the carry bag
The carry bag must be open while operating the
septic fluid suction pump.
Fig. 35: Operating in the carry bag
⌧
Hold the lid at the tabs.
⌧
Lift and swing the lid toward the rear.
The carry bag is now open.
⌧
Pull the two zippers near the tabs all the way
downward.
⌧
Swing the front panel down.
⌧
Remove the suction tube, fingertip and ex-
traction catheter from the side compartment
C
A U T I O N
!
Property damage due to overheating!
If the opening at the bottom of the carry bag were to be covered by film, paper or
the like, then the septic fluid suction pump would overheat during extraction.
The opening at the bottom of the carry bag must never be covered.
C
A U T I O N
!
Property damage due to overheating!
If placed on a soft surface (such as pillows or a mattress) the ventilation slots may
be covered and the septic fluid suction pump will overheat.
The septic fluid suction pump shall be upright and on a solid surface during oper-
ation.
W
A R N I N G
!
Infection hazard due to secretions spilling!
Transport the septic fluid suction pump only with the septic fluid jar empty.