User's Guide

CALIBRATION

Mount the gauge firmly with the loading shaft pointing downward. Go into the configu-
ration mode as described in the previous section and select the calibration sub-menu
by pressing ENTER three times when the display shows 'CAL'. After the display shows
'null' press ZERO, while insuring that there is no weight on the loading shaft other than
the weight of the required attachments (hooks, etc.). The next displayed prompt is
'SPAn' at which time apply the exact weight equal to the 

full capacity of  the gauge

in pounds

 and press ENTER. A successful calibration procedure is indicated by 'donE'

on the display. Press ENTER to save the new calibration data and to return to normal
operation. In some cases the display will show 'nnnn' or 'uuuu' to indicate excessive or
insufficient calibration weight. This can be caused by incorrect weights, tare weight of
over 10% of the full capacity of the gauge or an overloaded sensor. The calibration
procedure may be aborted  without changing the previous calibration information at
any time by pressing ESCAPE.

SPECIFICATIONS                              Section 11

General

Accuracy:

±

0.2% of full scale 

±

1 count

Tare capacity:

110% of capacity. Display shows "----" at 110%

Overload capacity:

150% of capacity. Display shows "----" at 110%

Sampling rate:

65 samples per second

Display update:

2.5-10 times per second in normal mode, depending on
filter settings. 65 times per second in peak mode

Display:

4-1/2-character LCD 0.3" [7.6 mm] high

Load  cell  deflection:

Maximum 0.010" [0.25 mm] at full scale

Outputs:

RS-232:

Baud rates between 300 and 9600

Mitutoyo:

Standard Mitutoyo SPC BCD output

Analog:

±

1 VDC 

±

0.25% FS

Connector:

9-pin D-type male

Power:

7.2 NiCd battery or included AC adapter/charger

Battery life:

8-10 hours per charge

Weight:

0.9 lbs [0.4 kg]

Capacity x graduation

BG012

0.12 x 0.00005 lbF

50 x 0.02 gF

0.5 x 0.0002 N

BG025

0.25 x 0.0001 lbF

100 x 0.05 gF

1 x 0.0005 N

BG05

0.5 x 0.0002 lbF

250 x 0.1 gF

2 x 0.001 N

BG2

2 x 0.001 lbF

1 x 0.0005 kgF

10 x 0.005 N

BG5

5 x 0.002 lbF

2 x 0.001 kgF

25 x 0.01 N

BG10

10 x 0.005 lbF

5 x 0.002 kgF

50 x 0.02 N

BG20

20 x 0.01 lbF

10 x 0.005 kgF

100 x 0.05 N

BG50

50 x 0.02 lbF

25 x 0.01 kgF

250 x 0.1 N

BG100

100 x 0.05 lbF

50 x 0.02 kgF

500 x 0.2 N

BG200

200 x 0.1 lbF

100 x 0.05 kgF

1000 x 0.5 N

BG500

500 x 0.2 lbF

250 x 0.1 kgF

2500 x 1 N

Section 10

Summary of Contents for BG012

Page 1: ...Series BG DIGITALFORCE GAUGES User s Guide...

Page 2: ...ECTION 4 6 EXTERNAL TRIGGER MODE SECTION 5 6 SET POINTS SECTION 6 7 AVERAGE MODE SECTION 7 7 GAUGE CONTROL LANGUAGE SECTION 8 8 OUTPUTS SECTION 9 10 RS 232 10 Mitutoyo BCD 10 Analog 10 I O connector p...

Page 3: ...dings DATA Initiates a data transmission sequence if equipped with the communication option MODE ADVANCE Switches the display between normal and peak modes of operation CLEAR ESCAPE Clears peak readin...

Page 4: ...a steady LO BAT indicating approximately one hour of charge remaining The second level is indicated by a flashing LO BAT At the third level the whole display except the LO BAT indicator will flash for...

Page 5: ...es for each reading FP 1 No filtering of peak readings FP 2 Average of 2 samples for each peak reading FP 4 Average of 4 samples for each peak reading FP 8 Average of 8 samples for each peak reading 2...

Page 6: ...itted 8 Every 8th sample transmitted 16 Every 16th sample transmitted 32 Every 32nd sample transmitted 64 Every 64th sample transmitted 128 Every 128th sample transmitted AoFF Automatic shutoff settin...

Page 7: ...or set to their minimum acceptable values for highest peak capture speed EXTERNAL TRIGGER MODE This mode of operation is useful for measuring electrical contact activation force as well as synchroniza...

Page 8: ...either case SP will appear on the display and the ENTER key may be used for re entering the set point change mode or the MODE key for proceeding with the normal operation of the gauge AVERAGE MODE Thi...

Page 9: ...Auto transmit every nth reading n 0 1 2 4 8 16 32 64 128 0 disabled LB Switch units to pounds KG Switch units to kilograms N Switch units to Newtons G Switch units grams ET Enable External trigger mo...

Page 10: ...ware version the last field shows the remaining battery power B0 full charge B3 minimum power All other fields show the status of settings and features using the same abbreviations as the commands to...

Page 11: ...ion to data collectors printers multiplexers or any other device capable of accepting Mitutoyo BCD data The transmission is initiated by the DATA key or by the receiving device see Section 3 for setti...

Page 12: ...ssing ESCAPE SPECIFICATIONS Section 11 General Accuracy 0 2 of full scale 1 count Tare capacity 110 of capacity Display shows at 110 Overload capacity 150 of capacity Display shows at 110 Sampling rat...

Page 13: ...nvironment or use other than intended will void this warranty This warranty is in lieu of all other warranties includ ing implied warranties of merchantability and fitness for an intended purpose In n...

Reviews: