Series BGI 

6

 

 

 

1-minute automatic shutoff 

 

 

5-minute automatic shutoff  

 10   

10-minute 

automatic shutoff 

 20   

20-minute 

automatic shutoff 

 30   

30-minute 

automatic shutoff 

 

init - Initial (default) settings 

 

LB 

 

Pounds as default units 

 

KG 

 

Kilograms as default units 

 

 

Grams as default units 

 

 

Newton's as default units 

 

OZIN 

 

Ounce-inches as default units 

 

LB IN 

 

Pound-inches as default units 

 

KGMM   

Kilogram-millimeters as default units 

 

NCM 

 

Newton-centimeters as default units  

 

 

NM 

 

Newton-meters as default units 

 

 

Average mode at turn on 

 

EtE 

 

External trigger mode at turn on 

 

TC 

or

 CCW 

Real time display at turn on 

 

PEAK  CCW 

Peak counter-clockwise mode at turn on 

 

PEAK CW 

Peak clockwise mode at turn on 

 

 PEAK 

 Peak 

tension 

mode at turn on 

 

PEAK C   

Peak compression mode at turn on 

 

A - Average mode settings 

 

A E 

 

Average mode enabled 

 

A d 

 

Average mode disabled

 

 dEL 

 Initial 

delay 

prompt 

 At 

 Average 

time 

prompt 

 trF 

 Trigger 

force/torque 

value 

prompt 

 

CAL - Calibration. See Section 10. 
 

4 FILTERS 

 

 

 

 

 

       

 
For maximum flexibility in noise suppression and peak capturing ability of the 
instrument there are two types of filters available to the user: analog and digital. 
The analog filter is a simple RC network with a cutoff frequency of 2.5 Hz and 
attenuation of 20 dB/decade. It can be either turned on or off. 
 
The digital filter utilizes the moving average technique in which consecutive 
readings are "pushed" through a buffer and the displayed reading is the aver-
age of the buffer contents. By varying the length of the buffer, a variable 
smoothing effect can be achieved. The BGI is equipped with a buffer which can 
hold up to eight readings. The number of readings to be averaged can be set to 
1,2,4 or 8. The selection of 1 will disable the filter since the average of a single 
value is the value itself. 
 

 

Summary of Contents for BGI

Page 1: ...DIGITAL FORCE TORQUE INDICATOR Model BGI User s Guide...

Page 2: ...ed For additional information or answers to your application questions contact us and our technical support and engineering teams will be eager to help you Thank you again for your purchase and happy...

Page 3: ...in parentheses For a detailed description of the secondary functions see Section 3 POWER ENTER Turns power on and off UNITS Selects units of measurement ZERO Zeroes any tare value up to the full capac...

Page 4: ...has several features with programmable options allowing many user specified choices In order to access the configuration menu perform the following 1 Turn off the indicator 2 Press and hold MODE 3 Tu...

Page 5: ...Set point outputs enabled bcd d Mitutoyo BCD output disabled bcd E Mitutoyo BCD output enabled nPOL Mitutoyo readings without polarity absolute value POL Mitutoyo readings with polarity positive for...

Page 6: ...gs A E Average mode enabled A d Average mode disabled dEL Initial delay prompt At Average time prompt trF Trigger force torque value prompt CAL Calibration See Section 10 4 FILTERS For maximum flexibi...

Page 7: ...trigger signal returns to its original state Please refer to the diagram for connection details and to Section 3 for configuration information 6 SET POINTS This feature is useful for tolerance checkin...

Page 8: ...The process of averaging begins as soon as the programmed trigger reading is reached and is indicated by a flashing A The conclusion of the test is indicated by an alternating A and the calculated val...

Page 9: ...e PKCW Peak Clockwise mode PKCCW Peak Counter Clockwise mode CLR Clear peaks start a new average or external trigger test Z Zero display and perform the CLR function Request the displayed reading C Re...

Page 10: ...ngs and features using the same abbreviations as the commands to set them Any detected errors are reported back by means of the following error codes 10 Illegal command 11 Not applicable e g SPHn comm...

Page 11: ...transmitted if DATA is pressed during the average computation while in the Average mode or during the input scanning in the External trigger mode Mitutoyo BCD This output is useful for connection to...

Page 12: ...eration In some cases the display will show nnnn or uuuu to indicate excessive or insufficient calibration weight This can be caused by incorrect weights tare weight of over 10 of the full capacity or...

Page 13: ...cts in workmanship and materials Mark 10 Corporation will at its option repair or replace or refund the purchased price of goods found to be defective This remedy shall be the buyer s sole and exclusi...

Page 14: ...standard line of products we can provide modifications and custom designs for OEM applications Our engineering team is eager to satisfy any special requirements Please contact us for further informati...

Reviews: