Series BGI
6
1
1-minute automatic shutoff
5
5-minute automatic shutoff
10
10-minute
automatic shutoff
20
20-minute
automatic shutoff
30
30-minute
automatic shutoff
init - Initial (default) settings
LB
Pounds as default units
KG
Kilograms as default units
G
Grams as default units
N
Newton's as default units
OZIN
Ounce-inches as default units
LB IN
Pound-inches as default units
KGMM
Kilogram-millimeters as default units
NCM
Newton-centimeters as default units
NM
Newton-meters as default units
A
Average mode at turn on
EtE
External trigger mode at turn on
TC
or
CCW
Real time display at turn on
PEAK CCW
Peak counter-clockwise mode at turn on
PEAK CW
Peak clockwise mode at turn on
PEAK
T
Peak
tension
mode at turn on
PEAK C
Peak compression mode at turn on
A - Average mode settings
A E
Average mode enabled
A d
Average mode disabled
dEL
Initial
delay
prompt
At
Average
time
prompt
trF
Trigger
force/torque
value
prompt
CAL - Calibration. See Section 10.
4 FILTERS
For maximum flexibility in noise suppression and peak capturing ability of the
instrument there are two types of filters available to the user: analog and digital.
The analog filter is a simple RC network with a cutoff frequency of 2.5 Hz and
attenuation of 20 dB/decade. It can be either turned on or off.
The digital filter utilizes the moving average technique in which consecutive
readings are "pushed" through a buffer and the displayed reading is the aver-
age of the buffer contents. By varying the length of the buffer, a variable
smoothing effect can be achieved. The BGI is equipped with a buffer which can
hold up to eight readings. The number of readings to be averaged can be set to
1,2,4 or 8. The selection of 1 will disable the filter since the average of a single
value is the value itself.