7.3 Set Points Outputs Schematic Diagram
7.4 Using Set Points to Control a Mark-10 Test Stand
When using set points to stop/cycle Mark-10 motorized test stands, the upper and lower set points must
be set to
opposite
measuring directions.
Both
set points must be set, even if the intended application is
to stop/cycle at only one of the set points. The opposite set point should be a value sufficiently large that it
does not get triggered during the course of the test.
For certain Mark-10 test stands, the upper and lower set point directions are reversed.
8 DATA MEMORY AND STATISTICS
Series 4 gauges have storage capacity of 50 data points. Readings may be stored, viewed, and output to
an external device. Individual, or all data points may be deleted. Statistics are calculated for the data in
memory.
To enable memory storage, select
DATA Key
from the menu, then scroll to
Memory Storage
and press
ENTER
. Then exit the menu. In the home screen, the data record number
00
appears below the primary
reading. Press
DATA
at any time to save the displayed reading. The record number will increment each
time
DATA
is pressed. If
DATA
is pressed when memory is full the message “MEMORY FULL” will be
flashed at the bottom of the display and a double audio tone will be sounded.
To view, edit, and output stored readings and statistics, select
Memory
from the menu. The screen
appears as follows:
MEMORY
View Data
View Statistics
Output Data
Output Statistics
Output Data & Stats
Clear All Data