Plug & Test
®
Smart Force & Torque Sensors
User’s Guide
19
14 SERIES R53 TORQUE SENSORS
14.1 Unpacking and Assembly
Carefully remove the sensor from the box. Install the four posts in the
desired positions on the sliders. If alternative gripping fixtures were
purchased, install them using the supplied hardware.
14.2 Overview
Designed for clockwise and counter-clockwise bottle cap torque testing.
Turning the knob moves the sliders along the length of the screw to engage
to disengage the sample. The posts or alternative gripping fixtures secure the sample while the cap is
twisted by hand. The sensor contains threaded holes on the underside of the base for bench mounting.
14.3 Specifications
Accuracy:
±0.5% of full scale
Safe overload:
MR53-10Z - MR53-50Z: 300% of full scale / MR53-12 - MR53-100: 150% of full scale
Operating temperature:
40ºF – 100ºF [5ºC – 38ºC]
Operating humidity:
96% max. (non-condensing)
Weight:
MR53-10Z - MR53-50Z: 2.2 lb [1.0 kg] / MR53-12 - MR53-100: 5.4 lb [2.5 kg]
14.4 Dimensions
(in[mm])
MR53-10Z - MR53-50Z:
MR53-12 - MR53-100:
14.5 Capacity x Resolution
Model
No.
With Model M7I / M5I indicator
With Model M3I indicator
ozFin
lbFin
lbFft
gFcm
kgFmm
Nmm
Ncm
Nm
ozFin
lbFin kgFmm Ncm
MR53-10Z
10 x 0.01
-
-
700 x 0.5
7 x 0.005
70 x 0.05
7 x 0.005
-
10 x 0.01
-
7 x 0.005
7 x 0.005
MR53-20Z
20 x 0.02
-
-
1400 x 1
14 x 0.01
140 x 0.1
14 x 0.01
-
20 x 0.02
-
14 x 0.01
14 x 0.01
MR53-50Z
50 x 0.05
-
-
3600 x 5
36 x 0.05
350 x 0.5
35 x 0.05
-
50 x 0.05
-
36 x 0.05
35 x 0.05
MR53-12
-
12 x 0.01
1 x 0.001
-
140 x 0.1
-
135 x 0.1
1.35 x 0.001
-
12 x 0.01
140 x 0.1
135 x 0.1
MR53-50
-
50 x 0.05
4 x 0.005
-
580 x 0.5
-
570 x 0.5
5.7 x 0.005
-
50 x 0.05
580 x 0.5
570 x 0.5
MR53-100
-
100 x 0.1
8 x 0.01
-
1150 x 1
-
1150 x 1
11.5 x 0.01
-
100 x 0.1
1150 x 1
1150 x 1
www.
GlobalTestSupply
.com
Find Quality Products Online at:
sales@GlobalTestSupply.com