5
4 SERIES R01 / R07 FORCE SENSORS
4.1 Unpacking and Assembly
Carefully remove the sensor from the box. No assembly is required.
4.2 Installing Series R07 to Models F755 / F755S / F1505 / F1505S
Test Frames
Install the sensor to the underside of the crosshead using the supplied
socket head screw and hardware. Plug the connector into the
receptacle in the top front surface of the crosshead, then fasten the
thumb screws into the crosshead, as shown below:
4.3 Overview
Tension and compression force may be applied to the surfaces with threaded holes. Attachments may be
threaded into these holes. These holes also allow for mounting as required.
4.4 Specifications
Accuracy:
±0.15% of full scale
Safe overload:
150% of full scale
Operating temperature:
40ºF – 100ºF [5ºC – 38ºC]
Operating humidity:
96% max. (non-condensing)
Weight:
1.9 to 3.0 lb [0.9 to 1.4 kg], depending on model
4.5 Dimensions
(in [mm])
Series R07 force sensors are identical to Series R01, except for two thumb screws in the Plug & Test
®
connector. These thumb screws are required for mounting to certain Series F test frames.