10.3 Command Set
Series TT01 testers will respond to the following ASCII commands via USB or RS-232:
?
Request the displayed reading
MEM
Transmit all stored readings
STA
Transmit
statistics
All commands must be terminated with a Carriage Return character or with a Carriage Return/Line Feed
combination. The tester’s responses are always terminated with a Carriage Return/Line Feed. Any
detected errors are reported back by means of error code *10 (illegal command).
10.4 Mitutoyo BCD settings
This output is useful for connection to data collectors, printers, multiplexers, or any other device capable
of accepting Mitutoyo BCD data. Individual data points may be transmitted by pressing
DATA
or by
requesting it from the Mitutoyo communication device (if available). Automatic output of individual data
points is also possible. Refer to the
Break Detection
sub-section for details. To enable Mitutoyo output,
select the desired format – either with polarity or without polarity. The screen appears as follows:
10.5 Break Detection
Three functions can be triggered automatically upon bottle cap removal (break detection), defined as a
50% drop in load from the peak. These functions automate the process of transmitting and saving
multiple individual test data, as well as zeroing the display. If beeps are enabled, an audible tone will
sound when the automatic output, automatic storage, and automatic zero functions have occurred. The
display will appear as follows:
Function Description
Auto Output
Automatically transmits the peak torque reading upon break detection. Press
ENTER
to access the
Auto Output Settings
sub-menu, where the output type
can be selected. Refer to Section 10.5.1 for details.
Auto Storage
Automatically stores the peak to memory upon break detection.
Auto Zero
Automatically zeroes the display following data transmission and/or storage.
Break Settings
Press
ENTER
to access the
Break Detection Settings
sub-menu to configure
the trigger for auto output and storage, and the delay to zero the display. Refer to
Section 10.5.2 for details.
BREAK DETECTION
+
Auto Output
Auto Storage
Auto Zero
+ Break Settings
Trig. Threshold
10
%
MITUTOYO BCD
*
Disabled
Ena w/o Polarity
Ena w/Polarity