Series TT01 Digital Cap Torque Testers
User’s Guide
18
If failed:
5. The following screen appears after the offsets have been calculated:
Attach weight fixtures (cable attachment, cable, hook, etc), as required. Do not yet attach any
weights or apply any calibration loads. Then press
ENTER
.
6. The display will appear as follows:
Optionally exercise the internal sensor several times (at full scale, if possible), then press
ENTER
.
7. The display will appear as follows:
Apply torque equal to the full scale of the instrument, then press
ENTER
.
CALIBRATION
CLOCKWISE
GAIN ADJUST
APPLY FULL SCALE LOAD
50.000 LBFIN +/-20%
THEN PRESS ENTER
CALIBRATION
CLOCKWISE
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
CLOCKWISE
Optionally exercise
load cell a few times.
THEN PRESS ENTER
CALIBRATION
OFFSET
Sen.Offset Adj.Failed
Ana.Offset Adj.Failed
CALIBRATION
OFFSET
Sen.Offset Adj.Passed
Ana.Offset Adj.Passed
CALIBRATION
OFFSET
Please wait…